Evidence map›Paper›PMID 33748025›Full record

ArticleHip & pelvis2021

Gross Trunnion Failure in the Bipolar Hemiarthroplasty; Raising Concern about Short Trunnion: A Case Report.

Kangbaek Kim, Junpyo Lee, Jeong Joon Yoo, Hee Joong Kim

Abstract readCase Reports
In one paragraph

Article in Hip & pelvis, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Review
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Kangbaek KimDepartment of Orthopaedic Surgery, Seoul National University Hospital, Seoul, Korea.
Junpyo LeeDepartment of Orthopaedic Surgery, Seoul National University Hospital, Seoul, Korea.
Jeong Joon YooDepartment of Orthopaedic Surgery, Seoul National University Hospital, Seoul, Korea.
Hee Joong KimDepartment of Orthopaedic Surgery, Seoul National University Hospital, Seoul, Korea.ORCID https://orcid.org/0000-0002-3994-5672

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

There have been some reports of gross trunnion failure (GTF) in total hip arthroplasty. Here, we report a case of GTF 19 years after bipolar hemiarthroplasty using a 28-mm head with a 14/16 taper bore. Compared to other GTF reports, the current case had some unusual aspects: bipolar hemiarthroplasty, 28-mm head, relatively late-onset, and no apparent findings of metallosis though a severe one was evident. A Computed tomography scout view provided valuable information in evaluating polyethylene, metal head, and neck inside the bipolar cup. The current report suggests a need for concerned regarding short trunnion length which may be associated with GTF.

Indexed as

ArthroplastyHipProsthesis failure

Identifiers

PMID33748025
PMCPMC7952271

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.