ArticleEJNMMI physics2021

Validation of CARE kV automated tube voltage selection for PET-CT: PET quantification and CT radiation dose reduction in phantoms.

Natalie A Bebbington, Troels Jørgensen, Erik Dupont, Mille A Micheelsen

Abstract read
In one paragraph

Article in EJNMMI physics, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed.

  1. Article
  2. Review
  3. Article
  4. EANM procedural recommendations for managing the paediatric patient in diagnostic nuclear medicine.European journal of nuclear medicine and molecular imaging · 2023
    Review
  5. Review
  6. Article
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Natalie A BebbingtonSiemens Healthcare A/S, Bredskifte Alle 15, 8210, Aarhus V, Denmark. natalie.bebbington@siemens-healthineers.com.ORCID http://orcid.org/0000-0003-4833-9030
Troels JørgensenDepartment of Nuclear Medicine, Zealand University Hospital, Køge, Denmark.
Erik DupontDepartment of Biomedical Engineering, Region Zealand, Roskilde, Denmark.
Mille A MicheelsenDepartment of Nuclear Medicine, Zealand University Hospital, Køge, Denmark.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundApplied tube voltage (kilovolts, kV) and tube current (milliampere seconds, mAs) affect CT radiation dose and image quality and should be optimised for the individual patient. CARE kV determines the kV and mAs providing the lowest dose to the patient, whilst maintaining user-defined reference image quality. Given that kV changes affect CT values which are used to obtain attenuation maps, the aim was to evaluate the effect of kV changes on PET quantification and CT radiation dose using phantoms.

methodFour phantoms ('Lungman', 'Lungman plus fat', 'Esser' and 'NEMA image quality' (NEMA IQ)) containing F-18 sources underwent 1 PET and 5 CT scans, with CARE kV on (automatic kV selection and mAs modulation) and in semi mode with specified tube voltages of 140, 120, 100 and 80 kV (mAs modulation only). A CARE kV image quality reference of 120 kV/50 mAs was used. Impact on PET quantification was determined by comparing measured activity concentrations for PET reconstructions from different CT scans with the reconstruction using the 120 kV reference, and dose (DLP, CTDI

resultsCARE kV-determined optimal tube voltage and CARE kV 'on' dose (DLP) savings compared with the 120 kV reference were: Lungman, 100 kV, 2.0%; Lungman plus fat, 120 kV, 0%; Esser, 100 kV, 9.3%; NEMA IQ, 100 kV, 3.4%. Using tube voltages in CARE kV 'semi' mode which were not advised by CARE kV 'on' resulted in dose increases ≤ 65% compared with the 120 kV reference (greatest difference Lungman plus fat, 80 kV). Clinically insignificant differences in PET activity quantification of up to 0.7% (Lungman, 100 kV, mean measured activity concentration) were observed when using the optimal tube voltage advised by CARE kV. Differences in PET quantification of up to 4.0% (Lungman, 140 kV, maximum measured activity concentration) were found over the full selection of tube voltages in semi mode, with the greatest differences seen at the most suboptimal kV for each phantom. However, most differences were within 1%.

conclusionsCARE kV on can provide CT radiation dose savings without concern over changes in PET quantification.

Indexed as

CARE kVdosekVoptimisationPET-CTquantification

Identifiers

PMID33743091
PMCPMC7981373