ArticlePhysical therapy2021

Smartphone Inclinometry Is a Valid and Reliable Tool for Measuring Frontal Plane Tibial Alignment in Healthy and Osteoarthritic Knees.

Calvin T F Tse, Jesse M Charlton, Jennifer Lam, Joanne Ho, Jessica Bears, Amanda Serek, Michael A Hunt

Open access · greenAbstract readValidation Study
In one paragraph

Article in Physical therapy, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed, 1 pooled it
0.5field-weighted citation impact, top 41% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 1 synthesis or guideline pooled it, 6 citations in OpenAlex.

  1. Pooled it
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 1 institution in 1 country.

Calvin T F TseGraduate Programs in Rehabilitation Sciences, University of British Columbia, Vancouver, British Columbia, Canada.
Jesse M CharltonGraduate Programs in Rehabilitation Sciences, University of British Columbia, Vancouver, British Columbia, Canada.
Jennifer LamDepartment of Physical Therapy, University of British Columbia, Vancouver, British Columbia, Canada.
Joanne HoDepartment of Physical Therapy, University of British Columbia, Vancouver, British Columbia, Canada.
Jessica BearsDepartment of Physical Therapy, University of British Columbia, Vancouver, British Columbia, Canada.
Amanda SerekDepartment of Physical Therapy, University of British Columbia, Vancouver, British Columbia, Canada.
Michael A HuntMotion Analysis and Biofeedback Laboratory, University of British Columbia, Vancouver, British Columbia, Canada.
University of British Columbia · CA

Funding

CIHR
6 · The paper itself

Abstract

objectiveFrontal plane knee alignment plays an integral role in tibiofemoral knee osteoarthritis development and progression. Accessible methods for obtaining direct or indirect measures of knee alignment may help inform clinical decision making when specialized equipment is unavailable. The present study evaluated the concurrent validity, as well as intersession (within-rater) and interrater (within-session) reliability of smartphone inclinometry for measuring static frontal plane tibial alignment-a known proxy of frontal plane knee alignment.

methodsTwenty healthy individuals and 38 patients with knee osteoarthritis were measured for frontal plane tibial alignment by a pair of raters using smartphone inclinometry, manual inclinometry, and 3-dimensional motion capture simultaneously. Healthy participants were measured on 2 separate days. Bland-Altman analysis, supplemented with intraclass correlation coefficient (ICC)(2,k), was used to assess concurrent validity. ICC(2,k), SEM, and minimum detectable change with 95% confidence limits (MDC95) were used to assess measurement reliability.

resultsCompared against motion capture, smartphone inclinometry measured frontal plane tibial alignment with a mean difference of 0.7 and 1.1 degrees (biased toward varus) for healthy participants and participants with knee osteoarthritis, respectively (ICC[2,k] ≥ 0.87). Smartphone inclinometry measurements demonstrated adequate intersession (within-rater) relative (ICC[2,k] = 0.91) and absolute (SEM = 0.7 degrees; MDC95 = 1.8 degrees) reliability, which outperformed manual inclinometry (ICC[2,k] = 0.85; SEM = 1.0 degrees; MDC95 = 2.6 degrees). Interrater (within-session) reliability of smartphone inclinometry was acceptable in both cohorts (ICC[2,k] = 0.93; SEM = 0.4 degrees to 1.2 degrees; MDC95 = 1.2 degrees to 3.2 degrees).

conclusionSmartphone inclinometry is sufficiently valid and reliable for measuring frontal plane tibial alignment in healthy individuals and patients with medial tibiofemoral knee osteoarthritis. IMPACT: Smartphones are readily accessible by clinicians and researchers. Our assessment of measurement validity and reliability supports the use of smartphone inclinometry as a clinically available tool to measure frontal plane tibial alignment without medical imaging or specialized equipment.

Indexed as

AdultBone MalalignmentFemaleHumansMaleMobile ApplicationsOsteoarthritis, KneeRange of Motion, ArticularReproducibility of ResultsSmartphoneTibiaYoung AdultAlignmentClinicalInclinometerOsteoarthritisSmartphone

Identifiers

PMID33710269
PMCPMC8248042
OpenAlexW3135892868

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.