Even the Simplest Devices May Malfunction: Split Septum Design Revisited.

Rotem Naftalovich, Steven Char, Andrew J Iskander, Daniel Naftalovich

Open access · bronzeAbstract read
In one paragraph

Article in Biomedical instrumentation & technology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact, top 96% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed, 0 citations in OpenAlex.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors at 2 institutions in 1 country.

Rotem Naftalovich
Steven Char
Andrew J Iskander
Daniel Naftalovich
Rutgers New Jersey Medical SchoolRutgers, The State University of New Jersey · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Split septum medical devices are used in tubing for intravenous (IV) fluid administration-an extremely common clinical task. These tubing caps contain a needleless, valveless system that allows fluid to flow directly through the lumen of the catheter but prevents backflow of fluid or blood when the tubing extension is not connected. We experienced complete failure of a needle-free connector extension set with a Luer-access split septum device in multiple patients due to the split septum remaining fused and essentially unsplit despite being connected on both ends. This led to an adverse event in a patient due to repeated unnecessary IV insertion attempts. This case shows how even the simplest of devices can malfunction and highlights the need for vigilance in clinical practice.

Indexed as

Catheters, IndwellingEquipment DesignHumansInfusions, Intravenous

Identifiers

PMID33690818
PMCPMC8641422
OpenAlexW3135165848

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.