ArticleInternational journal of computer assisted radiology and surgery2021
Efficient automated localization of ECoG electrodes in CT images via shape analysis.
Article in International journal of computer assisted radiology and surgery, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT04479410 (Efficient Automated Localization of ECoG Electrodes in CT Images Via Shape Analysis), which is not on this map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Efficient Automated Localization of ECoG Electrodes in CT Images Via Shape Analysis
Who cites it
5 citing papers in PubMed.
- BrainTRACE (Brain Tumor Registration and Cortical Electrocorticography): A novel tool for localizing electrocorticography electrodes in patients with brain tumors.Journal of neuroscience methods · 2025Article
- Anatomical registration of intracranial electrodes. Robust model-based localization and deformable smooth brain-shift compensation methods.Journal of neuroscience methods · 2024Review
- Advances in human intracranial electroencephalography research, guidelines and good practices.NeuroImage · 2022Review
- Modeling intracranial electrodes. A simulation platform for the evaluation of localization algorithms.Frontiers in neuroinformatics · 2022Article
- LeGUI: A Fast and Accurate Graphical User Interface for Automated Detection and Anatomical Localization of Intracranial Electrodes.Frontiers in neuroscience · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposePeople with drug-refractory epilepsy are potential candidates for surgery. In many cases, epileptogenic zone localization requires intracranial investigations, e.g., via ElectroCorticoGraphy (ECoG), which uses subdural electrodes to map eloquent areas of large cortical regions. Precise electrodes localization on cortical surface is mandatory to delineate the seizure onset zone. Simple thresholding operations performed on patients' computed tomography (CT) volumes recognize electrodes but also other metal objects (e.g., wires, stitches), which need to be manually removed. A new automated method based on shape analysis is proposed, which provides substantially improved performances in ECoG electrodes recognition.
methodsThe proposed method was retrospectively tested on 24 CT volumes of subjects with drug-refractory focal epilepsy, presenting a large number (> 1700) of round platinum electrodes. After CT volume thresholding, six geometric features of voxel clusters (volume, symmetry axes lengths, circularity and cylinder similarity) were used to recognize the actual electrodes among all metal objects via a Gaussian support vector machine (G-SVM). The proposed method was further tested on seven CT volumes from a public repository. Simultaneous recognition of depth and ECoG electrodes was also investigated on three additional CT volumes, containing penetrating depth electrodes.
resultsThe G-SVM provided a 99.74% mean classification accuracy across all 24 single-patient datasets, as well as on the combined dataset. High accuracies were obtained also on the CT volumes from public repository (98.27% across all patients, 99.68% on combined dataset). An overall accuracy of 99.34% was achieved for the recognition of depth and ECoG electrodes.
conclusionsThe proposed method accomplishes automated ECoG electrodes localization with unprecedented accuracy and can be easily implemented into existing software for preoperative analysis process. The preliminary yet surprisingly good results achieved for the simultaneous depth and ECoG electrodes recognition are encouraging. Ethical approval n°NCT04479410 by "IRCCS Neuromed" (Pozzilli, Italy), 30th July 2020.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials