ArticleBMC musculoskeletal disorders2021
Evaluation of an occipito-cervico fusion with a new implant design: a biomechanical study.
Article in BMC musculoskeletal disorders, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 0 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundA novel implant for occipitocervical fusion consisting of a median plate with an additional hook inserting in the foramen magnum was tested. Aim of this study was to test the stability of a new implant for occipitocervical fusion against the already available and employed median plate implant without hook. MATERIAL AND
method36 rigid polyurethane foams occipital artificial bones were used. The two occipital implants, namely the occipital plate with hook (Group 1) and the one without hook (Group 2), were applied to the artificial occiput trough three occipital screws and ensured into the experimental setup trough a crossbar. The test parameters were set using the testing machine software as follows: (1) test speed: 10 mm/ min, with 25 mm/ min maximum; (2) preload: 5 N; (3) force switch-off threshold: 90% force drop from F_max. Failure force and path were recorded. Failure force is defined as the maximum reaction force under which failure occurs (F_max), while failure path is the travel path during which failure occurs (dL).
resultsGroup 1 (plate with hook) showed a mean failure force of 459.3 ± 35.9 N and a mean failure path of 5.8 ± 0.3 mm Group 2 (plate without hook) showed a mean failure force of 323.9 ± 20.2 N and a mean failure path of 7.2 ± 0.4 mm. The Shapiro-Wilk test score was not significant (P > 0.1), assuming that data were normally distributed. Group 1 had a statistically significant greater F_max (+ 135.37; P > 0.0001) and less dL (- 1.52; P > 0.0001) compared to group 2.
conclusionsMedial plates with foramen magnum hooks showed to be more stable that plates without a hook. These new implants may represent a new tool in OCJ fixation, but further studies are required to investigate their behavior in an anatomical setting.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.