ArticleScientific reports2021
Performance of the hematology analyzer XN-31 prototype in the detection of Plasmodium infections in an endemic region of Colombia.
Article in Scientific reports, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
12 citing papers in PubMed, 1 synthesis or guideline pooled it, 26 citations in OpenAlex.
- Diagnostic role of Sysmex hematology analyzer in the detection of malaria: A systematic review and meta-analysis.PloS one · 2024Pooled it
- A comparison of the cost-effectiveness of the current standard of care and of the haematology analyser XN31-microscopy combination for diagnosing malaria in a nonendemic country.Memorias do Instituto Oswaldo Cruz · 2026Article
- Performance evaluation of the automated haematology analyzer XN-31 for malaria diagnosis in Plasmodium vivax-dominant regions of Thailand.Malaria journal · 2025Article
- The Diagnostic Performance of a Sysmex XN-31 Automated Malaria Analyzer vs. Expert Microscopy.International journal of laboratory hematology · 2025Article
- Autofluorescence of Red Blood Cells Infected withBiosensors · 2025Article
- Article
- Rapid Detection ofDiagnostics (Basel, Switzerland) · 2023Article
- Potential application of the haematology analyser XN-31 prototype for field malaria surveillance in Kenya.Malaria journal · 2022Article
- Clinical performance testing of the automated haematology analyzer XN-31 prototype using whole blood samples from patients with imported malaria in Japan.Malaria journal · 2022Article
- Fast detection and quantification of Plasmodium species infected erythrocytes in a non-endemic region by using the Sysmex XN-31 analyzer.Malaria journal · 2022Article
- Evaluation of the Sysmex XN-31 automated analyser for blood donor malaria screening at Malawi Blood Transfusion Services.Vox sanguinis · 2022Observational
- Diagnostic accuracy of fluorescence flow-cytometry technology using Sysmex XN-31 for imported malaria in a non-endemic setting.Parasite (Paris, France) · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors at 2 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
Early and accurate diagnosis is critical in reducing the morbidity and mortality associated with malaria. Microscopy (MI) is the current diagnostic gold standard in the field; however, it requires expert personnel, is time-consuming, and has limited sensitivity. Although rapid diagnostic tests for antigen detection (RDTs) are an alternative to diagnosis, they also have limited sensitivity and produce false positive results in detecting recent past infection. The automated hematology analyzer XN-31 prototype (XN-31p) (Sysmex Corporation, Kobe, Japan) is able to identify plasmodium-infected erythrocytes, count parasitemia and perform complete blood-cell counts within one minute. The performance of the XN-31p in diagnosing malaria was evaluated and compared with real-time polymerase chain reaction (qPCR), MI and RDT in an endemic area of Colombia where Plasmodium falciparum and Plasmodium vivax are present. Acute febrile patients were enrolled from July 2018 to April 2019 in Quibdó, Colombia. Malaria diagnoses were obtained from MI and RDT in the field and later confirmed by qPCR. Venous blood samples in EDTA were processed with an XN-31p in the field. Sensitivity, specificity, positive/negative predictive values, and the likelihood ratios of positive and negative tests were calculated with respect to the results from qPCR, MI and RDT. The intraclass correlation coefficient (ICC) and Bland-Altman plot were used to evaluate the concordance in the parasitemia with respect to MI. A total of 1,754 subjects were enrolled. The mean age was 27.0 years (IQR 14-44); 89.6% were Afro-Colombians, 94.3% lived in urban areas and 0.91% were pregnant. With respect to qPCR, the XN-31p showed a sensitivity of 90% (95% CI 87.24-92.34) and a specificity of 99.83% (95% CI 99.38-99.98) in detecting Plasmodium spp.; both parameters were equivalent to those for MI and RDT. Using MI as the reference, the XN-31p showed a sensitivity of 98.09% (95% CI 96.51-99.08), a specificity of 99.83% (95% CI 99.4-99.98), an ICC of 0.85 (95% CI 0.83-0.87) and an average difference of - 3096 parasites/µL when compared with thick-smear MI and an ICC of 0.98 (95% CI 0.97-0.98) and an average difference of - 0.0013% when compared with thin-smear MI. The XN-31p offers a rapid and accurate alternative method for diagnosing malaria in clinical laboratories in areas where P. falciparum and P. vivax cocirculate.