ArticleVeterinary ophthalmology2021
Evaluation of the effect of disposable tonometer cover brand on performance of Tono-Pen Vet in canine eyes.
Article in Veterinary ophthalmology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 0 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 1 institution in 1 country.
Funding
Abstract
PRIMARY
objectiveTo evaluate the effect of latex tip cover manufacturer on accuracy and repeatability of Tono-Pen Vet™ in canine eyes. ANIMAL STUDIED: Twelve enucleated globes from six dogs. PROCEDURES: The anterior chamber was cannulated and connected to a calibrated manometer. Intraocular pressure (IOP) measurements were obtained using the Tono-Pen Vet and TONOVET Plus at manometric IOP ranging from 5 to 80 mmHg. At each IOP, the Tono-Pen Vet was used with a new Ocu-Film™ latex tip cover (the only manufacturer-approved brand of cover) followed by a new Softips™ latex tip cover. For comparison, the TONOVET Plus was also used at each IOP with a new disposable rebound probe. Measured IOP values were analyzed by linear regression and intraclass correlation coefficient (ICC).
resultsTono-Pen Vet accuracy was unaffected by tip cover manufacturer or by frequent change in cover. Using ICC analysis, repeatability of measurements using either tonometer was good to excellent at physiologic IOP levels but variably decreased with both devices at supraphysiologic IOP.
conclusionsNeither tip cover manufacturer nor frequent changes in tip cover adversely affect Tono-Pen Vet accuracy. Measurement repeatability with Tono-Pen Vet and TONOVET Plus is widely variable at supraphysiologic IOP. Therefore, minor changes in IOP >25 mmHg should not be used to make clinical decisions without considering this variability.