Evidence map›Paper›PMID 33598521›Full record

ArticlePractical laboratory medicine2021

Improving quality of transcutaneous bilirubin measurements: Value of in-house developed quality control.

Heather Anne Paul, Brenda Joanne Adams, Allison Anne Venner

Abstract read
In one paragraph

Article in Practical laboratory medicine, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Article
  2. Article
  3. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Heather Anne PaulAlberta Precision Laboratories, 3535, Research Rd NW, Calgary, AB, T2L 2K8, Canada.
Brenda Joanne AdamsAlberta Precision Laboratories, 3535, Research Rd NW, Calgary, AB, T2L 2K8, Canada.
Allison Anne VennerAlberta Precision Laboratories, 3535, Research Rd NW, Calgary, AB, T2L 2K8, Canada.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectivesQuality assurance (QA) plays an integral role in Point of Care Testing (POCT) programs. Quality control (QC) is an important QA program component to ensure high quality results and enhanced patient care. The measurement of transcutaneous bilirubin (TcB) in the POCT setting is an essential part of newborn care in Alberta, Canada. However, there is currently no available commercial QC material for TcB meters. An in-house developed QC material has been in use within a single TcB POCT program within Alberta. The objective of this study was to determine the performance of this QC material by other POCT staff and clinical end-users to assess whether its use could be expanded. DESIGN AND

methodsTwo levels of QC material were measured by POCT staff and clinical end-users across 12 different sites using the Dräger Jaundice Meter JM-103® and JM-105® meters.

resultsThe performance of the QC material was acceptable when tested by POCT staff, was stable and reliable over time, and had an acceptable CV (≤8%). However, the data does not support use of the QC material by clinical end-users.

conclusionsThe use of the QC material could be expanded into other TcB settings for use by POCT staff. Additional training and experience with the QC material by end-users is needed to facilitate QC use in the clinical setting.

Indexed as

HyperbilirubinemiaOut-of-QC Range, OQRPOCTPoint of Care Testing, POCTQualityQuality Assurance, QAQuality controlQuality Control, QCTotal serum bilirubin, TSBTranscutaneous bilirubinTranscutaneous bilirubin, TcB