ArticleInternational urology and nephrology2021
Failure of ureteral stents subject to extrinsic ureteral obstruction and stent occlusions.
Article in International urology and nephrology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
10 citing papers in PubMed, 16 citations in OpenAlex.
- The interplay between bacterial biofilms, encrustation, and wall shear stress in ureteral stents: a review across scales.Frontiers in urology · 2023Review
- Magnetic resonance imaging of in vitro urine flow in single and tandem stented ureters subject to extrinsic ureteral obstruction.International journal of urology : official journal of the Japanese Urological Association · 2022Article
- Drainage of infected kidneys with ureteral stents: does size matter?World journal of urology · 2022Article
- Computational simulation of the flow dynamic field in a porous ureteric stent.Medical & biological engineering & computing · 2022Article
- Application, Advancement, and Complication of Ureteral Stent and Encrustation: A Major Complication.Cureus · 2022Review
- The accumulation of particles in ureteric stents is mediated by flow dynamics: Full-scale computational and experimental modeling of the occluded and unoccluded ureter.APL bioengineering · 2022Article
- Imaging and Chemical Analysis of External and Internal Ureteral Stent Encrustation.Research and reports in urology · 2022Article
- Response to: "Letter to the Editor, International Urology and Nephrology: in silico-in vitro-in vivo-can numerical simulations based on computational fluid dynamics (CFD) replace studies of the urinary tract?"International urology and nephrology · 2021Article
- Letter to the editor, international urology and nephrology-in silico-in vitro-in vivo: can numerical simulations based on computational fluid dynamics (CFD) replace studies of the urinary tract?International urology and nephrology · 2021Article
- Comparative study of renal drainage with different ureteral stents subject to extrinsic ureteral obstruction using an in vitro ureter-stent model.BMC urology · 2021Article
Corrections and comments
- Commented on by
Authors and funding
5 authors at 3 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo quantify the occurrence of stent failure and the dynamic behavior of urine flow in ureter-stent systems, including the relative flow in the ureter and stent lumina, subject to various degrees of ureter and stent blockage.
methodsNumerical simulations based on computational fluid dynamics (CFD) were used to quantify urine flow behavior in stented ureters, in the presence of extrinsic ureteral obstruction (EUO) and stent occlusions. Two stented ureter configurations were considered, one with circumferential occlusion of the ureter and the second with pressure on one side of the ureter wall. The pressure within the renal unit for different degrees of ureter closure and stent lumen occlusion was determined systematically. Onset of stent failure and the distribution of urine flow between stent and ureter lumina were determined.
resultsIn the case of EUO completely encircling the ureter, causing 100% obstruction of the ureter lumen, pressure in the renal unit is essentially unaffected until the stent lumen reaches ~ 90% occlusion, and fails only with > 95% occlusion. Occlusions of 50% in stent side holes in the vicinity of the EUO only alter local flow patterns but have no significant influence on renal unit pressure. For EUO deforming and compressing the ureter from one side, with ~ 50% reduction in ureter lumen, urine drainage proceeds with negligible increase in renal pressure even with 100% occlusion in the stent lumen.
conclusionCFD simulations show that stent failure under EUO tends to occur suddenly, only when both ureter and stent lumina become almost fully blocked.