Safety of mechanical lung vibrator and high-frequency chest wall oscillation in patients with cardiac implantable electronic device.

Hye Bin Gwag, Hyun Sung Joh, June Soo Kim, Kyoung-Min Park, Young Keun On, Seung-Jung Park

Open access · goldAbstract read
In one paragraph

Article in Clinical cardiology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
0.6field-weighted citation impact, top 31% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 6 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 2 institutions in 1 country.

Hye Bin GwagDivision of Cardiology, Department of Internal Medicine, Samsung Changwon Hospital, Sungkyunkwan University School of Medicine, Changwon, Republic of Korea.
Hyun Sung JohDivision of Cardiology, Department of Internal Medicine, Heart Vascular Stroke Institute, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
June Soo KimDivision of Cardiology, Department of Internal Medicine, Heart Vascular Stroke Institute, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
Kyoung-Min ParkDivision of Cardiology, Department of Internal Medicine, Heart Vascular Stroke Institute, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.ORCID https://orcid.org/0000-0002-8406-3142
Young Keun OnDivision of Cardiology, Department of Internal Medicine, Heart Vascular Stroke Institute, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
Seung-Jung ParkDivision of Cardiology, Department of Internal Medicine, Heart Vascular Stroke Institute, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.ORCID https://orcid.org/0000-0003-3288-0594
Samsung Medical Center · KRSamsung (South Korea) · KR

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundChest physiotherapy (CPT) is a non-pharmacological therapy to facilitate airway secretion removal. There have been concerns about potential electromagnetic interference (EMI) and lead integrity problems during the use of vibrating CPT devices in patients with cardiac implantable electronic devices (CIEDs). HYPOTHESIS: Two CPT devices can be used safely in patients with CIED.

methodsVolunteer patients with CIED underwent device interrogation to check lead integrity and device function before and after application of CPT devices. Mechanical lung vibrator and high-frequency chest wall oscillation (HFCWO) vests were used while monitoring surface electrocardiograms and intra-cardiac electrograms.

resultsWe prospectively enrolled 46 patients with CIEDs (25 pacemakers, 15 implantable cardioverter-defibrillators, and six cardiac resynchronization therapy-defibrillators). There was no noise detection or EMI during CPT in any patient. None of the patients showed clinically significant changes in lead integrity parameters. HFCWO inappropriately accelerated the pacing rate up to the maximal programmed value in five patients with pacemakers and two with cardiac resynchronization therapy-defibrillators.

conclusionCPT may be safely applied to patients with CIED without compromising lead integrity or device function, except for unwanted increase in pacing rate caused by misdetection of chest wall vibration as patients' activity while using HFCWO. Deactivation of the accelerometer-based activity sensor may be needed when HFCWO is planned for CPT.

Indexed as

Chest Wall OscillationDefibrillators, ImplantablePacemaker, ArtificialElectronicsHumansLungcardiac implantable electronic devicechest physiotherapydevice safety

Identifiers

PMID33590902
PMCPMC8027569
OpenAlexW3130218335

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.