Trial reportKnee surgery, sports traumatology, arthroscopy : official journal of the ESSKA2022
Patient-specific instruments do not show advantage over conventional instruments in unicompartmental knee arthroplasty at 2 year follow-up: a prospective, two-centre, randomised, double-blind, controlled trial.
Trial report in Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
6 citing papers in PubMed, 8 citations in OpenAlex.
- Patient-specific vs. Oxford microplasty instrumentation in unicompartmental knee arthroplasty: a randomized controlled trial.European journal of medical research · 2025Trial
- Patient-Specific Instrumentation Improves Technical Execution Accuracy After Artificial Intelligence-Assisted Preoperative Planning in Medial Unicompartmental Knee Arthroplasty.International orthopaedics · 2026Article
- Three dimensional printing in unicompartmental knee arthroplasty: current evidence and future directions.International orthopaedics · 2026Review
- Does patient-specific instrument or robot improve imaging and functional outcomes in unicompartmental knee arthroplasty? A bayesian analysis.Archives of orthopaedic and trauma surgery · 2024Article
- Article
- Patient Specific Instruments and Patient Individual Implants-A Narrative Review.Journal of personalized medicine · 2023Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors at 6 institutions in 3 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThe aim of this two-centre RCT was to compare pre- and post-operative radiological, clinical and functional outcomes between patient-specific instrumentation (PSI) and conventional instrumented (CI) unicompartmental knee arthroplasty (UKA). It was hypothesised that both alignment methods would have comparable post-operative radiological, clinical and functional outcomes.
methodsOne hundred and twenty patients were included, and randomly allocated to the PSI or the CI group. Outcome measures were peri-operative outcomes (operation time, length of hospital stay and intra-operative changes of implant size) and post-operative radiological outcomes including the alignment of the tibial and femoral component in the sagittal and frontal plane and the hip-knee-ankle-axis (HKA-axis), rate of adverse events (AEs) and patient-reported outcome measures (PROMs) pre-operatively and at 3, 12 and 24 months post-operatively.
resultsThere was a statistically significant difference (p < 0.05) in alignment of the femoral component in the frontal plane in favour of the CI method. No statistically significant differences were found for the peri-operative data or in the functional outcome at 2-year follow-up. In the PSI group, the approved implant size of the femoral component was correct in 98.2% of the cases and the tibial component was correct in 60.7% of the cases. There was a comparable rate of AEs: 5.1% in the CI and 5.4% in the PSI group.
conclusionThe PSI method did not show an advantage over CI in regard of positioning of the components, nor did it show an improvement in clinical or functional outcome. We conclude that the possible advantages of PSI do not outweigh the costs of the MRI scan and the manufacturing of the PSI. LEVEL OF EVIDENCE: Randomised controlled trial, level I.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.