Evaluation of Varian's SmartAdapt for clinical use in radiation therapy for patients with thoracic lesions.
Article in Journal of applied clinical medical physics, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionDeformable image registration (DIR) is a required tool in any adaptive radiotherapy program to help account for anatomical changes that occur during a multifraction treatment. SmartAdapt is a DIR tool from Varian incorporated within the eclipse treatment planning system, that can be used for contour propagation and transfer of PET, MRI, or computed tomography (CT) data. The purpose of this work is to evaluate the registration and contour propagation accuracy of SmartAdapt for thoracic CT studies using the guidelines from AAPM TG 132.
methodsTo evaluate the registration accuracy of SmartAdapt the mean target registration error (TRE) was measured for ten landmarked 4DCT images from the https://www.dir-labs.com/ which included 300 landmarks matching the inspiration and expiration phase images. To further characterize the registration accuracy, the magnitude of deformation for each 4DCT was measured and compared against the mean TRE for each study. Contour propagation accuracy was evaluated using 22 randomly selected lung cancer cases from our center where there was either a replan, or the patient was treated for a new lesion within the lung. Contours evaluated included the right and left lung, esophagus, spinal canal, heart and the GTV and the results were quantified using the DICE similarity coefficient.
resultsThe mean TRE from all ten cases was 1.89 mm, the maximum mean TRE per case was 3.8 mm from case #8, which also had the most landmark pairs with displacements >2 cm. For contour propagation accuracy, the DICE coefficient results for left lung, right lung, heart, esophagus, and spinal canal were 0.93, 0.94, 0.90, 0.61, and 0.82 respectively.
conclusionThe results from our study demonstrate that for thoracic images SmartAdapt in most cases will be accurate to below 2 mm in registration error unless there is deformation greater than 2 cm.