Evidence map›Paper›PMID 33565214›Full record

ArticleJournal of applied clinical medical physics2021

Automatic treatment planning for VMAT-based total body irradiation using Eclipse scripting.

Jose R Teruel, Sameer Taneja, Paulina E Galavis, K Sunshine Osterman, Allison McCarthy, Martha Malin, Naamit K Gerber, Christine Hitchen, David L Barbee

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 13 papers.

0numbers the graph read from it
0cells of the map it votes in
13citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

13 citing papers in PubMed.

  1. Trial
  2. VMAT-based total body irradiation on a conventional LINAC: workflow, procedure and preliminary results.Clinical & translational oncology : official publication of the Federation of Spanish Oncology Societies and of the National Cancer Institute of Mexico · 2026
    Review
  3. Article
  4. Article
  5. Article
  6. The Spinning Manny Indexed Overlay System for VMAT total body irradiation.Journal of applied clinical medical physics · 2026
    Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Automatic planning of the lower extremities for total marrow irradiation using volumetric modulated arc therapy.Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al] · 2023
    Article
  12. Review
  13. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

Jose R TeruelDepartment of Radiation Oncology, NYU Langone Health, New York, NY, USA.
Sameer TanejaDepartment of Radiation Oncology, NYU Langone Health, New York, NY, USA.
Paulina E GalavisDepartment of Radiation Oncology, NYU Langone Health, New York, NY, USA.
K Sunshine OstermanDepartment of Radiation Oncology, NYU Langone Health, New York, NY, USA.
Allison McCarthyDepartment of Radiation Oncology, NYU Langone Health, New York, NY, USA.
Martha MalinDepartment of Radiation Oncology, NYU Langone Health, New York, NY, USA.
Naamit K GerberDepartment of Radiation Oncology, NYU Langone Health, New York, NY, USA.
Christine HitchenDepartment of Radiation Oncology, NYU Langone Health, New York, NY, USA.
David L BarbeeDepartment of Radiation Oncology, NYU Langone Health, New York, NY, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The purpose of this work is to establish an automated approach for a multiple isocenter volumetric arc therapy (VMAT)-based TBI treatment planning approach. Five anonymized full-body CT imaging sets were used. A script was developed to automate and standardize the treatment planning process using the Varian Eclipse v15.6 Scripting API. The script generates two treatment plans: a head-first VMAT-based plan for upper body coverage using four isocenters and a total of eight full arcs; and a feet-first AP/PA plan with three isocenters that covers the lower extremities of the patient. PTV was the entire body cropped 5 mm from the patient surface and extended 3 mm into the lungs and kidneys. Two plans were generated for each case: one to a total dose of 1200 cGy in 8 fractions and a second one to a total dose of 1320 cGy in 8 fractions. Plans were calculated using the AAA algorithm and 6 MV photon energy. One plan was created and delivered to an anthropomorphic phantom containing 12 OSLDs for in-vivo dose verification. For the plans prescribed to 1200 cGy total dose the following dosimetric results were achieved: median PTV V100% = 94.5%; median PTV D98% = 89.9%; median lungs Dmean = 763 cGy; median left kidney Dmean = 1058 cGy; and median right kidney Dmean = 1051 cGy. For the plans prescribed to 1320 cGy total dose the following dosimetric results were achieved: median PTV V100% = 95.0%; median PTV D98% = 88.7%; median lungs Dmean = 798 cGy; median left kidney Dmean = 1059 cGy; and median right kidney Dmean = 1064 cGy. Maximum dose objective was met for all cases. The dose deviation between the treatment planning dose and the dose measured by the OSLDs was within ±4%. In summary, we have demonstrated that scripting can produce high-quality plans based on predefined dose objectives and can decrease planning time by automatic target and optimization contours generation, plan creation, field and isocenter placement, and optimization objectives setup.

Indexed as

Radiotherapy, Intensity-ModulatedWhole-Body IrradiationHumansRadiometryRadiotherapy DosageRadiotherapy Planning, Computer-Assistedadvanced treatment planningeclipse scriptingplan automationspecial procedurestotal body irradiation (TBI)

Identifiers

PMID33565214
PMCPMC7984467

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.