ArticleArthroplasty today2021
Variability of Cutting and Thermal Dynamics Between New and Used Acetabular Reamers During Total Hip Arthroplasty.
Article in Arthroplasty today, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 4 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundAseptic loosening of the acetabular component remains one of the leading causes of early failure of total hip arthroplasty. Poor apposition of bone onto the implant surface can be due to inaccurate reaming and osteonecrosis of the acetabular bone due to the heat generated while reaming.
methodsNew and used acetabular reamers were tested on an MTS system using a clinically relevant force of 87.6 N. A thermal profile and depth achieved by the reamers were analyzed and compared between the 2 cohorts. Heat generated and force required for the community used reamers to achieve the same depth as the new reamers were subsequently analyzed.
resultsThe new reamers achieved a depth 3.4 mm deeper than the community reamers (
conclusionsCommunity use of reamers will cause variations in depth of penetration and increased temperatures at a clinically generated force vs new reamers. When community reamers were forced to the same depths the new reamers achieved, a significantly greater amount of heat was generated, and an increased amount of time was needed, both of which are known risk factors for osteonecrosis.