Evidence map›Paper›PMID 33458354›Full record

ArticlePhysics and imaging in radiation oncology2020

Evaluation of automated pre-treatment and transit in-vivo dosimetry in radiotherapy using empirically determined parameters.

Evy Bossuyt, Reinhilde Weytjens, Daan Nevens, Sarah De Vos, Dirk Verellen

Open access · goldAbstract read
In one paragraph

Article in Physics and imaging in radiation oncology, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 31 papers.

0numbers the graph read from it
0cells of the map it votes in
31citing papers in PubMed
8.0field-weighted citation impact, top 2% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

31 citing papers in PubMed, 54 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Real-Time Dose-Guided Radiation Therapy.International journal of radiation oncology, biology, physics · 2025
    Review
  8. Article
  9. Article
  10. Article
  11. Article
  12. Article
  13. Article
  14. Article
  15. The role of EPID in vivo dosimetry in the risk management of stereotactic lung treatments.Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al] · 2023
    Article
  16. Article
  17. Review
  18. Selective de-implementation of routine in vivo dosimetry.Journal of applied clinical medical physics · 2023
    Review
  19. Article
  20. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 2 institutions in 1 country.

Evy BossuytIridium Kankernetwerk, Radiation Oncology Department, Antwerp, Belgium.
Reinhilde WeytjensIridium Kankernetwerk, Radiation Oncology Department, Antwerp, Belgium.
Daan NevensIridium Kankernetwerk, Radiation Oncology Department, Antwerp, Belgium.
Sarah De VosIridium Kankernetwerk, Radiation Oncology Department, Antwerp, Belgium.
Dirk VerellenIridium Kankernetwerk, Radiation Oncology Department, Antwerp, Belgium.
Iridium Kankernetwerk · BEUniversity of Antwerp · BE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

background and purposeFirst reports on clinical use of commercially automated systems for Electronic Portal Imaging Device (EPID)-based dosimetry in radiotherapy showed the capability to detect important changes in patient setup, anatomy and external device position. For this study, results for more than 3000 patients, for both pre-treatment verification and in-vivo transit dosimetry were analyzed. MATERIALS AND

methodsFor all Volumetric Modulated Arc Therapy (VMAT) plans, pre-treatment quality assurance (QA) with EPID images was performed. In-vivo dosimetry using transit EPID images was analyzed, including causes and actions for failed fractions for all patients receiving photon treatment (2018-2019). In total 3136 and 32,632 fractions were analyzed with pre-treatment and transit images respectively. Parameters for gamma analysis were empirically determined, balancing the rate between detection of clinically relevant problems and the number of false positive results.

resultsPre-treatment and in-vivo results depended on machine type. Causes for failed in-vivo analysis included deviations in patient positioning (32%) and anatomy change (28%). In addition, errors in planning, imaging, treatment delivery, simulation, breath hold and with immobilization devices were detected. Actions for failed fractions were mostly to repeat the measurement while taking extra care in positioning (54%) and to intensify imaging procedures (14%). Four percent initiated plan adjustments, showing the potential of the system as a basis for adaptive planning.

conclusionsEPID-based pre-treatment and in-vivo transit dosimetry using a commercially available automated system efficiently revealed a wide variety of deviations and showed potential to serve as a basis for adaptive planning.

Indexed as

In-vivoTransit dosimetry

Identifiers

PMID33458354
PMCPMC7807610
OpenAlexW3102011322

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked