Evidence map›Paper›PMID 33451019›Full record

ArticleSensors (Basel, Switzerland)2021

Achieving a Balanced Knee in Robotic TKA.

Alexander C Gordon, Michael A Conditt, Matthias A Verstraete

Open access · goldAbstract read
In one paragraph

Article in Sensors (Basel, Switzerland), 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 16 papers, 2 of them syntheses that pooled it.

0numbers the graph read from it
0cells of the map it votes in
16citing papers in PubMed, 2 pooled it
4.4field-weighted citation impact, top 5% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

16 citing papers in PubMed, 2 syntheses or guidelines pooled it, 29 citations in OpenAlex.

  1. Pooled it
  2. Pooled it
  3. Trial
  4. Trial
  5. Trial
  6. Article
  7. Article
  8. Article
  9. Review
  10. Article
  11. Article
  12. Article
  13. Article
  14. Observational
  15. Article
  16. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors at 2 institutions in 1 country.

Alexander C GordonIllinois Bone and Joint Institute, Des Plaines, IL 60016, USA.
Michael A CondittOrthoSensor Inc., Dania Beach, FL 33004, USA.ORCID 0000-0002-5966-4667
Matthias A VerstraeteOrthoSensor Inc., Dania Beach, FL 33004, USA.ORCID 0000-0002-4708-4235
OrthoSensor (United States) · USIllinois Bone and Joint Institute · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Total knee arthroplasty (TKA) surgery with manual instruments provides a quantitatively balanced knee in approximately 50% of cases. This study examined the effect of combining robotics technology with real-time intra-operative sensor feedback on the number of quantitatively balanced cases in a consecutive series of 200 robotic-assisted primary TKAs. The robotics platform was used to plan the implant component position using correctable poses in extension and a manual, centrally pivoting the balancer in flexion, prior to committing to the femoral cuts. During the initial trialing, the quantitative state of balance was assessed using an instrumented tibial tray that measured the intra-articular loads in the medial and lateral compartments. These sensor readings informed a number of surgical corrections, including bone recuts, soft-tissue corrections, and cement adjustments. During initial trialing, a quantitatively balanced knee was achieved in only 65% of cases. After performing the relevant soft-tissue corrections, bone recuts, and cement adjustments, 87% of cases ended balanced through the range of motion. Meanwhile, this resulted in a wide range of coronal alignment conditions, ranging from 6° valgus to 9° varus. It is therefore concluded that gaps derived from robotics navigation are not indicative for a quantitatively balanced knee, which was only consistently achieved when combining the robotics platform with real-time feedback from intra-operative load sensors.

Indexed as

Arthroplasty, Replacement, KneeOsteoarthritis, KneeAgedFemaleHumansKneeKnee JointMaleRange of Motion, Articularalignmentbalancingrobotic total knee arthroplasty

Identifiers

PMID33451019
PMCPMC7828562
OpenAlexW3118715762

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.