ArticleEuropean journal of orthopaedic surgery & traumatology : orthopedie traumatologie2021
Single lag screw and reverse distal femur locking compression plate for concurrent cervicotrochanteric and shaft fractures of the femur: biomechanical study validated with a clinical series.
Article in European journal of orthopaedic surgery & traumatology : orthopedie traumatologie, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed, 12 citations in OpenAlex.
- Biomechanical evaluation of a new intramedullary nail compared with proximal femoral nail antirotation and InterTAN for the management of femoral intertrochanteric fractures.Frontiers in bioengineering and biotechnology · 2024Article
- Biomechanical evaluation of two modified intramedullary fixation system for treating unstable femoral neck fractures: A finite element analysis.Frontiers in bioengineering and biotechnology · 2023Article
- Biomechanical evaluation of three implants for treating unstable femoral intertrochanteric fractures: finite element analysis in axial, bending and torsion loads.Frontiers in bioengineering and biotechnology · 2023Article
- Biomechanical study of extramedullary and intramedullary fixation in the treatment of unstable intertrochanteric reversed-tilt fractures of the femur.Annals of translational medicine · 2022Article
- A comparative Study of Novel Extramedullary Fixation and Dynamic Hip Screw in the Fixation of Intertrochanteric Fracture: A Finite-Element Analysis.Frontiers in surgery · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 3 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThe optimal surgical management of concurrent cervicotrochanteric and shaft fractures of the femur has not been consensual. The authors investigated the reliability of combined single lag screw and reverse distal femur locking compression plate (LCP-DF) by finite element (FE) study and retrospectively described the present technique for these dual fractures.
methodIntact femurs were derived from CT data, and the implant models were created by using CAD software. The fractured femur and implant models were virtually aligned based on the surgical techniques before converting to the FE model. In the FE model, applied boundary conditions included body weight, muscle forces, and constraint of the joints. Regarding clinical series, three patients with these dual fractures of the femur and 2 with cervicotrochanteric fractures with subtrochanteric extension were operated on by the proposed technique. The collected data include operative time, postoperative complications, union times, and clinical outcomes.
resultsEquivalent von Mises stress exhibited on dynamic hip screws with an anti-rotational screw was higher than the other techniques, close to the yield stress of the material. Multiple screw fixation produced better stability for transcervical fractures whereas the proposed technique of combined single lag screw and reverse LCP-DF provided better stability for intertrochanteric fractures. No significant difference in cortical bone stress was found between multiple screw construct and the proposed technique. The proposed technique presented a lower risk of secondary fractures, as the strain energy density (SED) in cancellous bone was lower than multiple screw construct. Regarding clinical series, all fractures were united with a mean union time of-16.1 weeks (range 12-20). There were no any postoperative complications. Regarding the Harris score, 1 was determined to be excellent value, and 4 to be good.
conclusionBy the FE results, a combination of a single lag screw and reverse LCP-DF is an effective technique for fixation of cervicotrochanteric fractures. Empowered by the clinical results, this proposed technique could be an alternative for concurrent cervicotrochanteric and shaft fractures of the femur especially when either single-system or dual-system devices seem not to be suitable.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials