Assessment of the image quality of virtual monochromatic spectral computed tomography images: a phantom study considering object contrast, radiation dose, and frequency characteristics.
Article in Radiological physics and technology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Evaluation of Image Quality and Detectability of Deep Learning Image Reconstruction (DLIR) Algorithm in Single- and Dual-energy CT.Journal of digital imaging · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
Fast kilovoltage (kVp)-switching technology cannot obtain conventional 120 kVp images; thus, 70 keV virtual monochromatic spectral computed tomography (CT) images (VMSI) are generally used. The contrast-to-noise ratio (CNR) is used to evaluate the image quality of VMSI; however, CNR does not include frequency characteristics. The present study aimed to investigate the evaluation methods of VMSI considering frequency characteristics by comparing the image quality of 70 keV VMSI with that of conventional 120 kVp images. The evaluated object contrasts were 70 and 300 Hounsfield units (HU). Scans used two radiation dose levels: low (LD) and standard (SD). The volume CT dose index of LD and SD was 4.8- and 12 mGy, respectively. Images were reconstructed by filtered back projection, evaluating CNR, noise power spectrum (NPS), task transfer function (TTF), and system performance (SP) function calculated as TTF