ArticleJournal of applied clinical medical physics2021
Initial testing of pegfilgrastim (Neulasta Onpro) on-body injector in multiple radiological imaging environments.
Article in Journal of applied clinical medical physics, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 3 citations in OpenAlex.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeAn increasing number of implantable or external devices can impact whether patients can receive radiological imaging examinations. This study examines and tests the Neulasta (pegfilgrastim) Onpro on-body injector in multiple imaging environments.
methodsThe injector was analyzed for four imaging modalities with testing protocols and strategies developed for each modality. In x-ray and computed tomography (CT), scans with much higher exposure than clinical protocols were performed with the device attached to an anthropomorphic phantom. The device was monitored until the completion of drug delivery. For magnetic resonance imaging (MRI), the device was assessed using a hand-held magnet and underwent the magnetically induced displacement testing in a 1.5T clinical MRI scanner room. For ultrasound, magnetic field changes were measured around an ultrasound scanner system with three transducers.
resultsFor x-ray and CT no sign of device error was identified during or after the high radiation exposure scans. Drug delivery was completed at expected timing with expected volume. For MRI the device showed significant attractive force towards the hand-held magnet and a 50-degree deflection angle at 50 cm from the opening of the scanner bore. No further assessment from the gradient or radiofrequency field was deemed necessary. For ultrasound the maximum magnetic field change from baseline was measured to be +11.7 μT in comparison to +74.2 μT at 4 inches from a working microwave.
conclusionsNo device performance issue was identified under the extreme test conditions in x-ray or CT. The device was found to be MR Unsafe. Magnetic field changes around an ultrasound system met the limitation set by manufacture. Patient ultrasound scanning is considered safe as long as the transducers do not inadvertently loosen the device.