Evaluation of technical performance of optical surface imaging system using conventional and novel stereotactic radiosurgery algorithms.

Hironori Kojima, Akihiro Takemura, Shogo Kurokawa, Shinichi Ueda, Kimiya Noto, Haruna Yokoyama, Shigeyuki Takamatsu

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers.

0numbers the graph read from it
0cells of the map it votes in
12citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

12 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Article
  12. Validation of CatalystIn vivo (Athens, Greece)
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Hironori KojimaDepartment of Radiology, Kanazawa University Hospital, Kanazawa, Ishikawa, Japan.ORCID https://orcid.org/0000-0001-9605-2742
Akihiro TakemuraFaculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, Kanazawa, Ishikawa, Japan.ORCID https://orcid.org/0000-0002-0103-3862
Shogo KurokawaDepartment of Radiation Technology, Shizuoka General Hospital, Shizuoka, Shizuoka, Japan.
Shinichi UedaDepartment of Radiology, Kanazawa University Hospital, Kanazawa, Ishikawa, Japan.
Kimiya NotoDepartment of Radiology, Kanazawa University Hospital, Kanazawa, Ishikawa, Japan.
Haruna YokoyamaDepartment of Radiology, Kanazawa University Hospital, Kanazawa, Ishikawa, Japan.
Shigeyuki TakamatsuDepartment of Radiation Therapy, Kanazawa University Hospital, Kanazawa, Ishikawa, Japan.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The Catalyst HD (C-RAD Positioning AB, Uppsala, Sweden) optical surface imaging (OSI) system is able to manage interfractional patient positioning, intrafractional motion monitoring, and non-contact respiratory gating without x-ray exposure for radiation therapy. In recent years, a novel high-precision surface registration algorithm for stereotactic radiosurgery (SRS algorithm) has been released. This study aimed to evaluate the technical performance of the OSI system using rigid phantoms, by comparing the conventional and SRS algorithms. To determine the system's technical performance, isocenter displacements were calculated by surface image registration via the OSI system using head, thorax, and pelvis rigid phantoms. The reproducibility of positioning was evaluated by the mean value calculated by repeating the registration 10 times, without moving each phantom. The accuracy of positioning was evaluated by the mean value of the residual error, where the 10 offset values given to each phantom were subtracted from the isocenter displacement values. The stability of motion monitoring was evaluated by measuring isocenter drift during 20 min and averaging it over 10 measurements. For the head phantom, all tests were compared with the mask types and algorithms. As a result, for all sites and both algorithms, the reproducibility, accuracy, and stability for translation and rotation were <0.1 mm and <0.1°, <1.0 mm and <1.0°, and <0.1 mm and <0.1°, respectively. In particular, the SRS algorithm had a small absolute error and standard deviation of calculated isocenter displacement, and a significantly higher reproducibility and accuracy than the conventional algorithm (P < 0.01). There was no difference in the stability between the algorithms (P = 0.0280). The SRS algorithm was found to be suitable for the treatment of rigid body sites with less deformation and small area, such as the head and face.

Indexed as

RadiosurgeryAlgorithmsHumansPhantoms, ImagingReproducibility of ResultsSwedenoptical surface imaging systempositioningregistrationstereotactic radiosurgery (SRS)surface-guided radiation therapy (SGRT)

Identifiers

PMID33369014
PMCPMC7882109

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.