ArticleJournal of clinical hypertension (Greenwich, Conn.)2021

The cuffless SOMNOtouch NIBP device shows poor agreement with a validated oscillometric device during 24-h ambulatory blood pressure monitoring.

Jakob Nyvad, Kent L Christensen, Niels Henrik Buus, Mark Reinhard

Open access · hybridAbstract read
In one paragraph

Article in Journal of clinical hypertension (Greenwich, Conn.), 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 14 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
14citing papers in PubMed, 1 pooled it
3.6field-weighted citation impact, top 5% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

14 citing papers in PubMed, 1 synthesis or guideline pooled it, 36 citations in OpenAlex.

  1. Pooled it
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Article
  12. Review
  13. Article
  14. Article
4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

4 authors at 1 institution in 1 country.

Jakob NyvadDepartment of Renal Medicine and Clinic of Hypertension, Aarhus University Hospital, Aarhus, Denmark.ORCID 0000-0002-5584-7407
Kent L ChristensenDepartment of Cardiology and Clinic of Hypertension, Aarhus University Hospital, Aarhus, Denmark.
Niels Henrik BuusDepartment of Renal Medicine and Clinic of Hypertension, Aarhus University Hospital, Aarhus, Denmark.
Mark ReinhardDepartment of Renal Medicine and Clinic of Hypertension, Aarhus University Hospital, Aarhus, Denmark.
Aarhus University Hospital · DK

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Repeated cuff-based blood pressure (BP) measurements may cause discomfort resulting in stress and erroneous recording values. SOMNOtouch NIBP is an alternative cuff-less BP measurement device that calculates changes in BP based on changes in pulse transit time (PTT) and a software algorithm. The device is calibrated with a single upper arm cuff-based BP measurement. We tested the device against a validated 24-h ambulatory BP monitoring (ABPM) device using both the previous (SomBP1) and the current software algorithm (SomBP2). In this study, 51 patients (mean age ± SD 61.5 ± 13.0 years) with essential hypertension underwent simultaneous 24-h ABPM with the SOMNOtouch NIBP on the left arm and a standard cuff-based oscillometric device on the right arm (OscBP). We found that mean daytime systolic BP (SBP) with OscBP was 140.8 ± 19.7 compared to 148.0 ± 25.2 (P = .008) and 146.9 ± 26.0 mmHg (P = .034) for SomBP1 and SomBP2, respectively. Nighttime SBP with OscBP was 129.5 ± 21.1 compared with 146.1 ± 25.8 (P < .0001) and 141.1 ± 27.4 mmHg (P = .001) for SomBP1 and SomBP2, respectively. Ninety-five% limits of agreement between OscBP and SomBP1 were ± 36.6 mmHg for daytime and ± 42.6 mmHg for nighttime SBP, respectively. Agreements were not improved with SomBP2. For SBP, a nocturnal dipping pattern was found in 33% of the study patients when measured with OscBP but only in 2% and 20% with SomBP1 and -2, respectively. This study demonstrates that BP values obtained with the cuff-less PTT-based SOMNOtouch device should be interpreted with caution as these may differ substantially from what would be obtained from a validated cuff-based BP device.

Indexed as

Blood Pressure Monitoring, AmbulatoryHypertensionBlood PressureBlood Pressure DeterminationHumansSphygmomanometersambulatory blood pressure/home blood pressure monitorclinical management of high blood pressure (HBP)hypertension - generalrisk assessmentvasoreactivity

Identifiers

PMID33350030
PMCPMC8030014
OpenAlexW3113212911

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.