In vitro investigation of the Flusso™ Bypass adapter efficiency upon ventilator circuit disconnect in a clinical simulated environment.
Article in Canadian journal of respiratory therapy : CJRT = Revue canadienne de la therapie respiratoire : RCTR, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
rationaleMechanically ventilated patients must be disconnected from the ventilator during intra-facility transfers. Intentional and accidental circuit disconnections represent a potential hazard to patients (sudden collapse and re-expansion of the alveoli) as well as to clinical staff (exposure to patient's unfiltered exhalation). Therefore, avoiding abrupt circuit disconnections could better protect the patient's health and reduce or eliminate contamination risks around clinical staff.
objectiveThe purpose of this in-vitro work was to investigate and evaluate the potential for environmental exposure of Nitric Oxide (NO, as an indicator of any contamination exposure) before and after implementing the novel Flusso™ Bypass adapter during the disconnect procedure of a mechanical ventilator system.
methodsA mechanical ventilator delivering NO was connected to a breathing simulator with and without the Flusso™ Bypass adapter. The ambient NO concentration was measured when the circuit was briefly disconnected (3 s) during inhalation and exhalation. Both volume and pressure ventilation modes were used. MEASUREMENTS AND MAIN
resultsDisconnecting the standard ventilator circuit (pressure-controlled mode) without the Flusso™ Bypass adapter produced higher NO escape to the surroundings (compared with the volume-controlled mode), leading to a longer NO dissipation time. No ambient NO traces were detected when the Flusso™ adapter was used.
conclusionThe usage of the Flusso™ adapter drastically decreases the unwanted exposure among clinical staff dealing with potentially hazardous airborne biological aerosols emanating from the circuit. Avoiding abrupt disconnection in the ventilator circuit could reduce lung injuries and alveolar over distension and collapse.