ArticleEuropean radiology experimental2018
Computed tomography image quality of aortic stents in patients with aortic coarctation: a multicentre evaluation.
Article in European radiology experimental, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed, 12 citations in OpenAlex.
- Assessing beam hardening artifacts in coronary stent imaging using different CT acquisition parameters on photon-counting detector computed tomography.The international journal of cardiovascular imaging · 2026Article
- Low-tube-potential ultra-high-resolution coronary CTA with photon-counting detector CT for stent evaluation: a comparative feasibility study.Japanese journal of radiology · 2025Article
- Dual-source photon-counting computed tomography for coronary in-stent observation: influence of heart rate and virtual monoenergetic image.The international journal of cardiovascular imaging · 2024Article
- Ultra-high-resolution photon-counting detector CT in evaluating coronary stent patency: a comparison to invasive coronary angiography.European radiology · 2024Article
- Non-Invasive Imaging Assessment in Patients with Aortic Coarctation: A Contemporary Review.Journal of clinical medicine · 2023Review
- Stent appearance in a novel silicon-based photon-counting CT prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating CT.European radiology experimental · 2023Article
- Computed Tomographic Evaluation of Congenital Left Ventricular Outflow Obstruction.Current cardiology reviews · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
12 authors at 3 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundStents are commonly used to treat aortic coarctation. The objective of this study was to evaluate the post-implantation computed tomography (CT) image quality of different stent types used to treat aortic coarctation.
methodsAdult and paediatric patients with stent-treated aortic coarctation who underwent contrast-enhanced CT were retrospectively included from three tertiary care centres. CT scans were subjectively scored for image quality using a 4-point scale (1 = unacceptable; 2 = poor; 3 = good; 4 = excellent). Furthermore, the amount of stent-induced blooming artefacts was measured as the percentage of the difference between outer and inner stent diameters over the outer stent diameter.
resultsA total of 35 children and 34 adults implanted with 71 stents of six different types were included. The most commonly used stent type was the Cheatham Platinum stent (52 stents, 73%). The subjective image quality of the Cheatham Platinum stents was moderate with a score of 2.0±0.8 (mean ± standard deviation) in children and 2.3±0.6 in adults. The image quality in patients with Formula stents was 2.3±1.2. The Cheatham Platinum stents induced 34-48% blooming, the Formula stents 44-55%. The image quality in patients with the less commonly used Atrium Advanta V12, IntraStent, AndraStent and Palmaz stents was scored 3 (good) to 4 (excellent) with less blooming. The electrocardiographic gating and tube voltage (kVp) did not affect image quality.
conclusionsThere is a substantial variation in CT image quality and blooming artefacts for different stent types used to treat aortic coarctation.