ArticleEJNMMI physics2020
Direct comparison of brain [
Article in EJNMMI physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
11 citing papers in PubMed.
- Myocardial flow reserve estimation from standardized uptake values ofAnnals of nuclear medicine · 2026Article
- Assessing small-lesion detectability and acquisition time optimisation in silicon-detector-Based PET: a phantom study.EJNMMI physics · 2025Article
- Quantitative validation of data-driven motion correction for brain PET using phantom with motion generator system.EJNMMI physics · 2025Article
- Elucidation of FDG-PET Imaging Characteristics that Reflect the Heterogeneity within Tumors Due to Variability in Metabolic Activity: A Phantom Study.World journal of nuclear medicine · 2025Article
- Optimization of penalization function in Bayesian penalized likelihood reconstruction algorithm for [Physical and engineering sciences in medicine · 2024Article
- Phantom and clinical evaluation of the Bayesian penalised likelihood reconstruction algorithm Q.Clear without PSF correction in amyloid PET images.EJNMMI physics · 2024Article
- Comparative study of physiological FDG uptake in small structures between silicon photomultiplier-based PET and conventional PET.Annals of nuclear medicine · 2024Article
- High-Resolution Silicon Photomultiplier Time-of-Flight Dedicated Head PET System for Clinical Brain Studies.Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2023Article
- Differential Diagnosis of Alzheimer Disease vs. Mild Cognitive Impairment Based on Left Temporal Lateral Lobe Hypomethabolism onDiagnostics (Basel, Switzerland) · 2022Article
- Diagnostic performance of a novel high-resolution dedicated axillary PET system in the assessment of regional nodal spread of disease in early breast cancer.Quantitative imaging in medicine and surgery · 2022Article
- EANM procedure guidelines for brain PET imaging using [European journal of nuclear medicine and molecular imaging · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundSilicon photomultiplier-positron emission tomography (SiPM-PET) has better sensitivity, spatial resolution, and timing resolution than photomultiplier tube (PMT)-PET. The present study aimed to clarify the advantages of SiPM-PET in
methodsContrast was calculated from images acquired from a Hoffman 3D brain phantom, and image noise and uniformity were calculated from images acquired from a pool phantom using SiPM- and PMT-PET. Sequential PMT-PET and SiPM-PET [
resultsImage contrast, image noise, and uniformity in SiPM-PET changed 19.2, 3.5, and - 40.0% from PMT-PET, respectively. These physical indices of both PET scanners satisfied the criteria for acceptable image quality published by the Japanese Society of Nuclear Medicine of contrast > 55%, CV ≤ 15%, and SD ≤ 0.0249, respectively. Contrast was 70.0% for SiPM-PET without TOF and 59.5% for PMT-PET without TOF. The TOF improved contrast by 3.5% in SiPM-PET. The SUV
conclusionsThe improved spatial resolution and sensitivity of SiPM-PET contributed to better image contrast and uniformity in brain [