Evidence map›Paper›PMID 33211375›Full record

ArticleJournal of applied clinical medical physics2020

Characterization of radiotherapy component impact on MR imaging quality for an MRgRT system.

Benjamin C Lewis, Bruce Gu, Ryan Klett, Rajiv Lotey, Olga L Green, Taeho Kim

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Benjamin C LewisDepartment of Radiation Oncology, Washington University School of Medicine, St Louis, MO, USA.ORCID https://orcid.org/0000-0003-1638-833X
Bruce GuDepartment of Radiation Oncology, Washington University School of Medicine, St Louis, MO, USA.
Ryan KlettBJC HealthCare, St Louis, MO, USA.
Rajiv LoteyViewRay Inc, Oakwood Village, OH, USA.
Olga L GreenDepartment of Radiation Oncology, Washington University School of Medicine, St Louis, MO, USA.
Taeho KimDepartment of Radiation Oncology, Washington University School of Medicine, St Louis, MO, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Radiotherapy components of an magnetic resonnace-guided radiotherapy (MRgRT) system can alter the magnetic fields, causing spatial distortion and image deformation, altering imaging and radiation isocenter coincidence and the accuracy of dose calculations. This work presents a characterization of radiotherapy component impact on MR imaging quality in terms of imaging isocenter variation and spatial integrity changes on a 0.35T MRgRT system, pre- and postupgrade of the system. The impact of gantry position, MLC field size, and treatment table power state on imaging isocenter and spatial integrity were investigated. A spatial integrity phantom was used for all tests. Images were acquired for gantry angles 0-330° at 30° increments to assess the impact of gantry position. For MLC and table power state tests all images were acquired at the home gantry position (330°). MLC field sizes ranged from 1.66 to 27.4 cm edge length square fields. Imaging isocenter shift caused by gantry position was reduced from 1.7 mm at gantry 150° preupgrade to 0.9 mm at gantry 120° postupgrade. Maximum spatial integrity errors were 0.5 mm or less pre- and postupgrade for all gantry angles, MLC field sizes, and treatment table power states. However, when the treatment table was powered on, there was significant reduction in SNR. This study showed that gantry position can impact imaging isocenter, but spatial integrity errors were not dependent on gantry position, MLC field size, or treatment table power state. Significant isocenter variation, while reduced postupgrade, is cause for further investigation.

Indexed as

Magnetic Resonance ImagingParticle AcceleratorsHumansMagnetic FieldsPhantoms, ImagingMRgRTMRI isocenter variationMR-LINAC

Identifiers

PMID33211375
PMCPMC7769410

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.