ArticleMedical physics2021
Comparison of automated beam hardening correction (ABHC) algorithms for myocardial perfusion imaging using computed tomography.
Article in Medical physics, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed.
- Cardiac CT Perfusion Imaging of Pericoronary Adipose Tissue (PCAT) Highlighting Potential Confounds in CTA Analysis.Journal of clinical medicine · 2025Article
- Quantitative cardiac CT perfusion: physiologically-inspired model and identifying microvascular disease from discordant CTA CAD-RADS.Frontiers in cardiovascular medicine · 2025Article
- Reproducibility of a single-volume dynamic CT myocardial blood flow measurement technique: validation in a swine model.European radiology experimental · 2024Article
- Angle Dependence of Electrode Lead-Related Artifacts in Single- and Dual-Energy Cardiac ECG-Gated CT Scanning-A Phantom Study.Journal of clinical medicine · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
Abstract
purposeMyocardial perfusion imaging using computed tomography (MPI-CT) and coronary CT angiography (CTA) have the potential to make CT an ideal noninvasive imaging gatekeeper exam for invasive coronary angiography. However, beam hardening can prevent accurate blood flow estimation in dynamic MPI-CT and can create artifacts that resemble flow deficits in single-shot MPI-CT. In this work, we compare four automatic beam hardening correction algorithms (ABHCs) applied to CT images, for their ability to produce accurate single images of contrast and accurate MPI flow maps using images from conventional CT systems, without energy sensitivity.
methodsPreviously, we reported a method, herein called ABHC-1, where we iteratively optimized a cost function sensitive to beam hardening artifacts in MPI-CT images and used a low order polynomial correction on projections of segmentation-processed CT images. Here, we report results from two new algorithms with higher order polynomial corrections, ABHC-2 and ABHC-3 (with three and seven free parameters, respectively), having potentially better correction but likely reduced estimability. Additionally, we compared results to an algorithm reported by others in the literature (ABHC-NH). Comparisons were made on a digital static phantom with simulated water, bone, and iodine regions; on a digital dynamic anthropomorphic phantom, with simulated blood flow; and on preclinical porcine experiments. We obtained CT images on a prototype spectral detector CT (Philips Healthcare) scanner that provided both conventional and virtual keV images, allowing us to quantitatively compare corrected CT images to virtual keV images. To test these methods' parameter optimization sensitivity to noise, we evaluated results on images obtained using different mAs.
resultsIn images of the static phantom, ABHC-2 reduced beam hardening artifacts better than our previous ABHC-1 algorithm, giving artifacts smaller than 1.8 HU, even in the presence of high noise which should affect parameter optimization. Taken together, the quality of static phantom results ordered ABHC-2> ABHC-3> ABHC-1>> ABHC-NH. In an anthropomorphic MPI-CT simulator with homogeneous myocardial blood flow of 100 ml⋅min
conclusionThe automated algorithm can be used with different parametric BH correction models. ABHC-2 improved MPI-CT blood flow estimation as compared to other approaches and was robust to noisy images. In simulation and preclinical experiments, ABHC-2 gave results approaching gold standard 70 keV measurements.