Evidence map›Paper›PMID 33185754›Full record

ArticleEJNMMI physics2020

Halo artifacts of indwelling urinary catheter by inaccurate scatter correction in

Keiichi Magota, Naoto Numata, Daiki Shinyama, Junya Katahata, Yamato Munakata, Piotr J Maniawski, Kentaro Kobayashi, Osamu Manabe, Kenji Hirata, Ukihide Tateishi and 2 more

Abstract read
In one paragraph

Article in EJNMMI physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Article
  2. Article
  3. Differential privacy preserved federated transfer learning for multi-institutionalEuropean journal of nuclear medicine and molecular imaging · 2023
    Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

12 authors.

Keiichi MagotaDivision of Medical Imaging and Technology, Hokkaido University Hospital, Sapporo, Japan.
Naoto NumataDivision of Medical Imaging and Technology, Hokkaido University Hospital, Sapporo, Japan.
Daiki ShinyamaPhilips Japan, Tokyo, Japan.
Junya KatahataDivision of Medical Imaging and Technology, Hokkaido University Hospital, Sapporo, Japan.
Yamato MunakataDivision of Medical Imaging and Technology, Hokkaido University Hospital, Sapporo, Japan.
Piotr J ManiawskiPhilips Healthcare, Cleveland, Ohio, USA.
Kentaro KobayashiDepartment of Diagnostic Imaging, Hokkaido University Graduate School of Medicine, Kita 15 Nishi 7, Kita-ku, Sapporo, Hokkaido, 060-8638, Japan.
Osamu ManabeDepartment of Diagnostic Imaging, Hokkaido University Graduate School of Medicine, Kita 15 Nishi 7, Kita-ku, Sapporo, Hokkaido, 060-8638, Japan.
Kenji HirataDepartment of Diagnostic Imaging, Hokkaido University Graduate School of Medicine, Kita 15 Nishi 7, Kita-ku, Sapporo, Hokkaido, 060-8638, Japan.
Ukihide TateishiDepartment of Diagnostic Radiology and Nuclear Medicine, Graduate School, Tokyo Medical and Dental University, Tokyo, Japan.
Kohsuke KudoDepartment of Diagnostic Imaging, Hokkaido University Graduate School of Medicine, Kita 15 Nishi 7, Kita-ku, Sapporo, Hokkaido, 060-8638, Japan.
Tohru ShigaDepartment of Diagnostic Imaging, Hokkaido University Graduate School of Medicine, Kita 15 Nishi 7, Kita-ku, Sapporo, Hokkaido, 060-8638, Japan. tshiga@med.hokudai.ac.jp.ORCID http://orcid.org/0000-0002-3238-3507

Funding

The Ministry of Education,Culture,Sports,Science and Technology 18K18412The Ministry of Education,Culture,Sports,Science and Technology 19K17127
6 · The paper itself

Abstract

backgroundHalo artifacts from urinary catheters can occur due to inaccurate scatter correction, and the artifacts affect the tumor visibility in

methodsWe conducted patient and phantom studies. (1) We retrospectively reviewed the cases of patients who had undergone

results(1) There were 61 patients with urinary catheters; in five (8.2%), halo artifacts were observed in the TFS 4-mm PET images. (2) The phantom study clearly reproduced the halo artifacts in the TFS 4-mm PET images. The halo artifacts were generated when urine moved in the interval between the CT and PET imaging, and when the urinary catheter was placed in a circular shape. The SUVs for the TFS 4-mm and TFS-2mm PET images were underestimated at the halo-artifact regions, whereas the SUVs for the MCS 4-mm PET images were close to the true values. (3) The halo artifacts disappeared in the TFS 2-mm PET images in 4/5 patients but not 1/5 patient, whereas the halo artifacts were completely absent in the MCS 4-mm PET images in 5/5 patients.

conclusionsThese data suggest that halo artifacts are caused if the PET images do not correspond to the physical material in the μ-maps, which induces the scatter correction error. With the MC-SSS, it was possible to accurately estimate the scatter without generating halo artifacts.

Indexed as

18F-FDGHalo artifactPET/CTScatter correctionUrinary catheter

Identifiers

PMID33185754
PMCPMC7666262