ArticleJournal of applied clinical medical physics2020
A phantom study to evaluate three different registration platform of 3D/3D, 2D/3D, and 3D surface match with 6D alignment for precise image-guided radiotherapy.
Article in Journal of applied clinical medical physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed.
- Intra-fractional lung tumor motion monitoring using arbitrary gantry angles during radiotherapy treatment.Zeitschrift fur medizinische Physik · 2025Article
- Computer-aided robotics for applications in fracture reduction surgery: Advances, challenges, and opportunities.iScience · 2025Review
- Markerless motion tracking with simultaneous MV and kV imaging in spine SBRT treatment-a feasibility study.Physics in medicine and biology · 2023Article
- Implementation of advanced radiotherapy techniques: stereotactic body radiotherapy (SBRT) for oligometastatic patients with lung metastasis - a single institution experience.Medicine and pharmacy reports · 2022Article
- Commissioning of optical surface imaging systems for cranial frameless stereotactic radiosurgery.Journal of applied clinical medical physics · 2021Article
- Accuracy of surface-guided patient setup for conventional radiotherapy of brain and nasopharynx cancer.Journal of applied clinical medical physics · 2021Article
- Commissioning and Quality Assurance of CatalystJournal of medical physicsArticle
- Accumulated Dose Deviation of Rotational and Residual Setup Errors on Nasopharyngeal Carcinoma Using MIM Treated by Helical Tomotherapy.Technology in cancer research & treatmentArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
Abstract
purposeTo evaluate two three-dimensional (3D)/3D registration platforms, one two-dimensional (2D)/3D registration method, and one 3D surface registration method (3DS). These three technologies are available to perform six-dimensional (6D) registrations for image-guided radiotherapy treatment.
methodsFiducial markers were asymmetrically placed on the surfaces of an anthropomorphic head phantom (n = 13) and a body phantom (n = 8), respectively. The point match (PM) solution to the six-dimensional (6D) transformation between the two image sets [planning computed tomography (CT) and cone beam CT (CBCT)] was determined through least-square fitting of the fiducial positions using singular value decomposition (SVD). The transformation result from SVD was verified and was used as the gold standard to evaluate the 6D accuracy of 3D/3D registration in Varian's platform (3D3DV), 3D/3D and 2D/3D registration in the BrainLab ExacTrac system (3D3DE and 2D3D), as well as 3DS in the AlignRT system. Image registration accuracy from each method was quantitatively evaluated by root mean square of target registration error (rmsTRE) on fiducial markers and by isocenter registration error (IRE). The Wilcoxon signed-rank test was utilized to compare the difference of each registration method with PM. A P < 0.05 was considered significant.
resultsrmsTRE was in the range of 0.4 mm/0.7 mm (cranial/body), 0.5 mm/1 mm, 1.0 mm/1.5 mm, and 1.0 mm/1.2 mm for PM, 3D3D, 2D3D, and 3DS, respectively. Comparing to PM, the mean errors of IRE were 0.3 mm/1 mm for 3D3D, 0.5 mm/1.4 mm for 2D3D, and 1.6 mm/1.35 mm for 3DS for the cranial and body phantoms respectively. Both of 3D3D and 2D3D methods differed significantly in the roll direction as compared to the PM method for the cranial phantom. The 3DS method was significantly different from the PM method in all three translation dimensions for both the cranial (P = 0.003-P = 0.03) and body (P < 0.001-P = 0.008) phantoms.
conclusion3D3D using CBCT had the best image registration accuracy among all the tested methods. 2D3D method was slightly inferior to the 3D3D method but was still acceptable as a treatment position verification device. 3DS is comparable to 2D3D technique and could be a substitute for X-ray or CBCT for pretreatment verification for treatment of anatomical sites that are rigid.