Total skin electron beam therapy rationalization and utility of

Sarah Misson-Yates, Marium Naeem, Isabel Palmer, Eleanor Holden, Owen Hedley, Mark McGovern, Stephen Morris, Anthony G Greener

Open access · goldAbstract read
In one paragraph

Article in BJR open, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
0.6field-weighted citation impact, top 29% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 6 citations in OpenAlex.

  1. Total skin electron beam therapy.Frontiers in oncology · 2025
    Article
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 1 institution in 1 country.

Sarah Misson-YatesDepartment of Medical Physics, Guy's and St Thomas' NHS Foundation Trust, SE1 9RT, London, UK.ORCID 0000-0002-3471-6088
Marium NaeemDepartment of Medical Physics, Guy's and St Thomas' NHS Foundation Trust, SE1 9RT, London, UK.
Isabel PalmerDepartment of Medical Physics, Guy's and St Thomas' NHS Foundation Trust, SE1 9RT, London, UK.
Eleanor HoldenDepartment of Medical Physics, Guy's and St Thomas' NHS Foundation Trust, SE1 9RT, London, UK.
Owen HedleyDepartment of Medical Physics, Guy's and St Thomas' NHS Foundation Trust, SE1 9RT, London, UK.
Mark McGovernDepartment of Medical Physics, Guy's and St Thomas' NHS Foundation Trust, SE1 9RT, London, UK.
Stephen MorrisDepartment of Clinical Oncology, Guy's and St Thomas' NHS Foundation Trust, SE1 9RT, London, UK.
Anthony G GreenerDepartment of Medical Physics, Guy's and St Thomas' NHS Foundation Trust, SE1 9RT, London, UK.
Guy's and St Thomas' NHS Foundation Trust · GB

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectiveThis paper reports on the rationalization of a substantial pool of

methodsThermoluminescent dosimeter IVD data from 149 patients were analyzed comparing the population mean and standard deviation for each site. The number of sites required to confirm the prescribed dose were reviewed considering both dosimetric and clinical relevance. The reduced sites were then used to assess the continued dosimetric accuracy on new equipment and the results were compared statistically using the Mann-Witney test.

resultsThe trunk dose measurement points were reduced from nine to six and five extra trunk sites were identified and reviewed clinically prior to removal.Following change in manufacturer the trunk dose points showed no statistically significant change and confirmed that patients had received within 1.3% of the intended mean trunk dose using both delivery methods.A statistically significant change in 4 out of the 13 extra trunk sites was seen following the move to the new centre. However, all but one site showed a change of less than 1 standard deviation.

conclusionThe total number of measurement points per patient were reduced from 27 to 19 which constituted a 25% saving in preparation and read out.Accurate delivery of prescribed dose was confirmed following measurement point reduction for treatments delivered on linacs from two different manufacturers. ADVANCES IN KNOWLEDGE: Proven methodology for rationalization of IVD measurements for TSEBT.

Identifiers

PMID33178940
PMCPMC7592482
OpenAlexW2962024381

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.