ArticleJournal of applied clinical medical physics2020
The quantification and potential impact of dark current on treatments with an MR-guided radiotherapy (MRgRT) system.
Article in Journal of applied clinical medical physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 1 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeDark current radiation produced during linac beam-hold has the potential to lead to unplanned dose delivered to the patient. With the increased usage of motion management and step-and-shoot IMRT deliveries for MR-guided systems leading to increased beam-hold time, it is necessary to consider the impact of dark current radiation on patient treatments.
methodsThe relative dose rate due to dark current for the ViewRay MRIdian linac was measured longitudinally over 15 months (June 2018-August 2019). Ion chamber measurements were acquired with the linac in the beam-hold state and the beam-on state, with the ratio representing the relative dark current dose rate. The potential contribution of the dark current dose to the overall prescription was retrospectively analyzed for 972 fractions from 83 patients over the same time period. The amount of time spent in the beam-hold state was combined with the monthly measured relative dark current dose rate to estimate the dark current dose contribution.
resultsThe relative dark current dose rate compared to the beam-on dose rate was 0.12% ± 0.027%. In a near worst-case estimation, the dark current dose contribution accounted for 0.90% ± 0.67% of the prescription dose across all fractions (3.61% maximum). Gantry and MLC motion between segments accounted for 87% of the dark current contribution, with the remaining 13% attributable to gating during segment delivery. The largest dark current contributions were associated with plans delivering a small dose per treatment segment.
conclusionsThe dark current associated with new clinical treatment units should be considered prior to treatment delivery to ensure it will not lead to dosimetric inaccuracies. For the MRIdian linac system investigated in this work, the contribution from dark current remained relatively low, though users should be cognizant of the larger potential dosimetric contribution for plans with small doses per segment.