Evidence map›Paper›PMID 33078562›Full record

ArticleJournal of applied clinical medical physics2020

CBCT image quality QA: Establishing a quantitative program.

Sameer Taneja, David L Barbee, Anthony J Rea, Martha Malin

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers.

0numbers the graph read from it
0cells of the map it votes in
10citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

10 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Sameer TanejaDepartment of Radiation Oncology, New York University Langone Medical Center, New York, NY, USA.
David L BarbeeDepartment of Radiation Oncology, New York University Langone Medical Center, New York, NY, USA.
Anthony J ReaDepartment of Radiation Oncology, New York University Langone Medical Center, New York, NY, USA.
Martha MalinDepartment of Radiation Oncology, New York University Langone Medical Center, New York, NY, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeRoutine quality assurance (QA) of cone-beam computed tomography (CBCT) scans used for image-guided radiotherapy is prescribed by the American Association of Physicists in Medicine Task Group (TG)-142 report. For CBCT image quality, TG-142 recommends using clinically established baseline values as QA tolerances. This work examined how image quality parameters vary both across machines of the same model and across different CBCT techniques. Additionally, this work investigated how image quality values are affected by imager recalibration and repeated exposures during routine QA.

methodsCone-beam computed tomography scans of the Catphan 604 phantom were taken on four TrueBeam® and one Edge™ linear accelerator using four manufacturer-provided techniques. TG-142 image quality parameters were calculated for each CBCT scan using SunCHECK Machine™. The variability of each parameter with machine and technique was evaluated using a two-way ANOVA test on a dataset consisting of 200 CBCT scans. The impact of imager calibration on image quality parameters was examined for a subset of three machines using an unpaired Student's t-test. The effect of artifacts appearing on CBCTs taken in rapid succession was characterized and an approach to reduce their appearance was evaluated. Additionally, a set of baselines and tolerances for all image quality metrics was presented.

resultsAll imaging parameters except geometric distortion varied with technique (P < 0.05) and all imaging parameters except slice thickness varied with machine (P < 0.05). Imager calibration can change the expected value of all imaging parameters, though it does not consistently do so. While changes are statistically significant, they may not be clinically significant. Finally, rapid acquisition of CBCT scans can introduce image artifacts that degrade CBCT uniformity.

conclusionsThis work characterized the variability of acquired CBCT data across machines and CBCT techniques along with the impact of imager calibration and rapid CBCT acquisition on image quality.

Indexed as

Radiotherapy, Image-GuidedSpiral Cone-Beam Computed TomographyCone-Beam Computed TomographyHumansParticle AcceleratorsPhantoms, Imagingcone-beam computed tomographyimage qualityinstitutional baselineslinear accelerator quality assurance

Identifiers

PMID33078562
PMCPMC7701111

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.