ArticleJournal of applied clinical medical physics2020
Dosimetric and mechanical equivalency of Varian TrueBeam linear accelerators.
Article in Journal of applied clinical medical physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed, 15 citations in OpenAlex.
- Accelerated and standardized commissioning of a robotic RadioSurgery system using a hybrid golden beam data model.Journal of applied clinical medical physics · 2026Article
- Developing reference plans for evaluating global clinical trials credentialing and PSQA systems.Journal of applied clinical medical physics · 2025Article
- 3D gamma analysis between treatment plans for nominally beam-matched medical linear accelerators using PyMedPhys.Precision radiation oncology · 2024Article
- A radiotherapy community data-driven approach to determine which complexity metrics best predict the impact of atypical TPS beam modeling on clinical dose calculation accuracy.Journal of applied clinical medical physics · 2024Article
- The assessment of the clinical impact of using a single set of radiotherapy planning data for two kilovoltage therapy units.Physical and engineering sciences in medicine · 2024Article
- Matched linac stereotactic radiotherapy: An assessment of delivery similarity and distributive patient-specific quality assurance feasibility.Journal of applied clinical medical physics · 2022Article
- Golden beam data provided by linear accelerator manufacturers should be used in the commissioning of treatment planning systems.Physical and engineering sciences in medicine · 2022Article
- Transferability of patients for radiation treatment between unmatched machines.Journal of applied clinical medical physics · 2022Article
- Cross-modality Dosimetry Audit in Head-and-neck Radiotherapy: A Phantom-based Approach for Photon and Proton Beams.Journal of medical physicsArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo investigate and improve the level of equivalency of Varian TrueBeam linear accelerators (linacs) in energy-, dosimetric leaf gap- (DLG) and jaw calibration.
methodsEight linacs with four photon energies: 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV, and three electron energies (on two linacs): 6, 9, and 12 MeV were commisioned and beam-matched. Initially, symmetry of lateral profiles was calibrated for maximum field size. Energy-matching was then performed for photons by adjusting diagonal profiles at maximum field size and depth of maximum dose to coincide with the reference linac, and for electrons by matching the range at percentage depth of ionization of 90%, 80%, and 50%. Calibration of DLG was performed for 6 MV and evaluated among the linacs. The relationship between DLG and the Gap value was investigated. A method using electronic portal imaging device (EPID) was developed and implemented for jaw calibration.
resultsSymmetry calibration for photons (electrons) was within 1% (0.7%), further improving the vendor's acceptance criteria. Photon and electron energy-matching was within 0.5% and 0.1 mm, respectively. Calibration of DLG was within 0.032 mm among the linacs and utilizing the relationship between DLG and the Gap value resulted in an empirical calibration method which was implemented to simplify DLG adjustment. Using EPID-based method of calibration, evaluation of the jaw-positioning among the linacs for 30 cm × 30 cm field size was within 0.4 mm and in the junction area within 0.2 mm. Dose delivery error of VMAT-plans were at least 99.2% gamma pass rate (1%, 1 mm).
conclusionsHigh level of equivalency, beyond clinically accepted criteria, of TrueBeam linacs could be achieved which reduced dose delivery systematic errors and increased confidence in interchanging patients among linacs.