The MOVE-C Cervical Artificial Disc - Design, Materials, Mechanical Safety.

Annette Kienle, Nicolas Graf, Carina Krais, Hans-Joachim Wilke

Open access · goldAbstract read
In one paragraph

Article in Medical devices (Auckland, N.Z.), 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
0.7field-weighted citation impact, top 29% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 5 citations in OpenAlex.

  1. Article
  2. Article
  3. [Three-dimensional finite element study on combined proximal and distal knee extension rearrangement for recurrent patellar dislocation].Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chinese journal of reparative and reconstructive surgery · 2022
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Annette KienleSpineServ GmbH & Co. KG, Ulm 89077, Germany.ORCID 0000-0003-1778-2670
Nicolas GrafSpineServ GmbH & Co. KG, Ulm 89077, Germany.
Carina KraisSpineServ GmbH & Co. KG, Ulm 89077, Germany.
Hans-Joachim WilkeInstitute of Orthopaedic Research and Biomechanics, Ulm University Medical Centre, Ulm 89081, Germany.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThere are various cervical disc prostheses on the market today. They can be subdivided into implants with a ball-and-socket design and implants with a flexible core, which is captured between the implant endplates and sealed using various sheaths. Implants with an articulating surface are mostly metal-on-metal or metal-on-UHMWPE designs and, thus, do not allow for axial damping. The aim of this study is to provide mechanical safety and performance data of the MOVE-C cervical disc prosthesis which combines both an articulating surface and a flexible core. MATERIALS AND

methodsMOVE-C consists of a cranial and caudal metal plate made of TiAl6V4. The cranial plate is TiNbN coated on its articulating surface. The caudal plate has a fixed polycarbonate-urethane (PCU) core. The TiNbN coating is meant to optimize the wear behavior of the titanium endplate, whereas the PCU core is meant to allow for a reversible axial deformation, a pre-defined neutral zone and a progressive load-deformation curve in all planes.

resultsVarious standard testing procedures (for example, ISO 18192-1 and ASTM F2364) and non-standard mechanical tests were carried out to prove the implant's mechanical safety. Due to the new implant design, wear and creep testing was deemed most important. The wear rate for the PCU was in maximum 1.54 mg per million cycles. This value was within the range of the UHMWPE wear rates reported for other cervical disc prostheses (0.53 to 2.59 mg/million cycles). Also in the creep-relaxation test, a qualitatively physiological behavior was shown with a certain amount of remaining deformation but no failure.

conclusionThe mechanical safety of the MOVE-C cervical disc prosthesis was shown to be comparable to other cervical disc prostheses. Since PCU wear particles were elsewhere shown to be less bioactive than cross-linked UHMWPE particles, wear-related failure in vivo may be less frequent compared to other prostheses. This, however, will have to be shown in further studies.

Indexed as

cervical spinecreepdegenerative disc diseasedisc arthroplastypolycarbonate-urethanewear