Evidence map›Paper›PMID 33048364›Full record

ArticleActa paediatrica (Oslo, Norway : 1992)2021

Optimisation of clinical workflow and monitor settings safely reduces alarms in the NICU.

Gabriele Varisco, Heidi van de Mortel, Laura Cabrera-Quiros, Louis Atallah, Dirk Hueske-Kraus, Xi Long, Eduardus Je Cottaar, Zhuozhao Zhan, Peter Andriessen, Carola van Pul

Abstract read
In one paragraph

Article in Acta paediatrica (Oslo, Norway : 1992), 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers.

0numbers the graph read from it
0cells of the map it votes in
10citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

10 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Observational
  6. Article
  7. Article
  8. Review
  9. Article
  10. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors.

Gabriele VariscoApplied Physics, Eindhoven University of Technology, Eindhoven, The Netherlands.ORCID 0000-0002-9107-0420
Heidi van de MortelPediatrics, Máxima Medical Center, Veldhoven, The Netherlands.
Laura Cabrera-QuirosApplied Physics, Eindhoven University of Technology, Eindhoven, The Netherlands.
Louis AtallahPhilips Healthcare, Andover, MA, USA.
Dirk Hueske-KrausPhilips Healthcare, Böblingen, Germany.
Xi LongPhilips Research, Eindhoven, The Netherlands.
Eduardus Je CottaarApplied Physics, Eindhoven University of Technology, Eindhoven, The Netherlands.
Zhuozhao ZhanMathematics and Computer Science, Eindhoven University of Technology, Eindhoven, The Netherlands.
Peter AndriessenApplied Physics, Eindhoven University of Technology, Eindhoven, The Netherlands.
Carola van PulApplied Physics, Eindhoven University of Technology, Eindhoven, The Netherlands.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

aimTo address alarm fatigue, a new alarm management system which ensures a quicker delivery of alarms together with waveform information on nurses' handheld devices was implemented and settings optimised. The effects of this clinical implementation on alarm rates and nurses' responsiveness were measured in an 18-bed single family rooms neonatal intensive care unit (NICU).

methodsThe technical implementation of the alarm management system was followed by clinical workflow optimisation. Alarms and vital parameters from October 2017 to December 2019 were analysed. Measures included monitoring alarms, nurses' response to alarms and time spent by patients in different saturation ranges. A survey among nurses was performed to evaluate changes in alarm rate and use of protocols.

resultsA significant reduction of monitoring alarms per patient days was detected after the optimisation phase (in particular for SpO2 ≤ 80%, P < .001). More time was spent by infants within the optimal peripheral oxygen saturation range (88% < SpO2 < 95%, P < .001). Results from the surveys showed that false alarms are less likely to cause an inappropriate response after the optimisation phase.

conclusionThe implementation of an alarm management solution and an optimisation programme can safely reduce the alarm burden inside of the NICU environment.

Indexed as

Clinical AlarmsIntensive Care Units, NeonatalHumansInfantInfant, NewbornMonitoring, PhysiologicSurveys and QuestionnairesWorkflowalarmalarm managementmonitoringneonatal intensive care unitsafety

Identifiers

PMID33048364
PMCPMC7983880

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.