ArticleOphthalmic research2022
Ex vivo Comparison of Intraocular Pressure Fluctuation during Pars Plana Vitrectomy Performed Using 25- and 27-Gauge Systems.
Article in Ophthalmic research, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 4 citations in OpenAlex.
- Risk factors for iatrogenic retinal breaks during induction of posterior vitreous detachment in macula surgery.International ophthalmology · 2026Article
- Updates on Treatment Modalities for Primary Rhegmatogenous Retinal Detachment Repair.Diagnostics (Basel, Switzerland) · 2024Review
- Mechanical Property Comparison of 23-, 25-, and 27-Gauge Trocars across Multiple Pressures and Vitrectomy Systems.Ophthalmology scienceArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionThe purpose of this study was to compare intraoperative intraocular pressure fluctuation using different aspiration systems and 25- and 27-gauge vitreous surgery probes.
methodsEx vivo, pars plana, 25- and 27-gauge vitreous surgery was performed on 4 porcine eyes, and IOP fluctuations were evaluated. We performed 3-port vitrectomy using the Constellation® Vision or the EVA® Phaco-Vitrectomy system. Each 20-s experiment was conducted 5 times for each set of conditions, each with the same substituted balanced salt solution. Real-time intraoperative intraocular pressure measurement was performed at the distal end of the infusion tube. Intraocular pressure was measured during core vitrectomy, core vitrectomy with fluid aspiration, peripheral vitreous shaving with scleral indentation, and fluid-gas exchange. The Mann-Whitney U test was used to evaluate statistical significance.
resultsMean ± standard deviation intraoperative intraocular pressure fluctuation during 25- and 27-gauge core vitrectomy was 15.9 ± 1.6 and 11.9 ± 1.4 mm Hg, respectively (p < 0.05), using the Constellation system; 23.2 ± 1.4 and 14.1 ± 0.7 mm Hg, respectively (p < 0.001), using the EVA vacuum mode; and 15.0 ± 0.5 and 11.5 ± 1.4 mm Hg, respectively (p < 0.05), using the EVA flow mode. The smallest intraoperative intraocular pressure fluctuations during core vitrectomy with fluid aspiration, peripheral vitreous shaving with scleral indentation, and fluid-gas exchange were all achieved using the 27-gauge EVA flow mode; these values were 14.2 ± 0.4, 35.7 ± 0.9, and 6.4 ± 0.2 mm Hg, respectively.
conclusionRegardless of the aspiration system, intraoperative intraocular pressure fluctuation was lower during 27-gauge than during 25-gauge vitrectomy. The 27-gauge EVA flow mode produced optimal intraoperative intraocular pressure stability.