Evidence map›Paper›PMID 33005395›Full record

ArticleBone & joint research2020

A higher degree of polyethylene irradiation is associated with a reduced risk of revision for aseptic loosening in total hip arthroplasties using cemented acetabular components: an analysis of 290,770 cases from the National Joint Registry of England, Wales, Northern Island and the Isle of Man.

Edward T Davis, Joseph Pagkalos, Branko Kopjar

Open access · goldAbstract read
In one paragraph

Article in Bone & joint research, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed, 1 pooled it
1.6field-weighted citation impact, top 16% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed, 1 synthesis or guideline pooled it, 10 citations in OpenAlex.

  1. Pooled it
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors at 2 institutions in 2 countries.

Edward T DavisRoyal Orthopaedic Hospital NHS Foundation Trust, Birmingham, UK.
Joseph PagkalosRoyal Orthopaedic Hospital NHS Foundation Trust, Birmingham, UK.ORCID 0000-0002-9890-9831
Branko KopjarRoyal Orthopaedic Hospital NHS Foundation Trust, Birmingham, UK.
Royal Orthopaedic Hospital · GBUniversity of Washington · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

aimsTo investigate the effect of polyethylene manufacturing characteristics and irradiation dose on the survival of cemented and reverse hybrid total hip arthroplasties (THAs).

methodsIn this registry study, data from the National Joint Registry of England, Wales, Northern Ireland and the Isle of Man (NJR) were linked with manufacturing data supplied by manufacturers. The primary endpoint was revision of any component. Cox proportional hazard regression was a primary analytic approach adjusting for competing risk of death, patient characteristics, head composition, and stem fixation.

resultsA total of 290,770 primary THAs were successfully linked with manufacturing characteristics. Overall 4,708 revisions were analyzed, 1,260 of which were due to aseptic loosening. Total radiation dose was identified as a risk factor and included in the Cox model. For statistical modelling of aseptic loosening, THAs were grouped into three categories: G1 (no radiation); G2 ( > 0 to < 5 Mrad); and G3 ( ≥ 5 Mrad). G1 had the worst survivorship. The Cox regression hazard ratio for revision due to aseptic loosening for G2 was 0.7 (95% confidence interval (CI) 0.58 to 0.83), and for G3 0.4 (95% CI 0.30 to 0.53). Male sex and uncemented stem fixation were associated with higher risk of revision and ceramic heads with lower risk.

conclusionPolyethylene irradiation was associated with reduced risk of revision for aseptic loosening. Radiation doses of ≥ 5 Mrad were associated with a further reduction in risk.Cite this article:

Indexed as

Aseptic looseningCrosslinkingIrradiationPolyethyleneRegistryTotal hip arthroplasty

Identifiers

PMID33005395
PMCPMC7502255
OpenAlexW3083864282

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.