Evidence map›Paper›PMID 32989859›Full record

ArticleMagnetic resonance in medicine2021

Prospective motion correction for diffusion weighted EPI of the brain using an optical markerless tracker.

Johan Berglund, Adam van Niekerk, Henric Rydén, Tim Sprenger, Enrico Avventi, Ola Norbeck, Stefan L Glimberg, Oline V Olesen, Stefan Skare

Open access · hybridAbstract read
In one paragraph

Article in Magnetic resonance in medicine, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers.

0numbers the graph read from it
0cells of the map it votes in
12citing papers in PubMed
1.7field-weighted citation impact, top 13% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

12 citing papers in PubMed, 23 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Review
  9. External Hardware and Sensors, for Improved MRI.Journal of magnetic resonance imaging : JMRI · 2023
    Review
  10. Review
  11. Article
  12. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors at 4 institutions in 3 countries.

Johan BerglundDepartment of Clinical Neuroscience, Karolinska Institutet, Stockholm, Sweden.ORCID 0000-0002-0853-9305
Adam van NiekerkDepartment of Clinical Neuroscience, Karolinska Institutet, Stockholm, Sweden.ORCID 0000-0001-9731-6930
Henric RydénDepartment of Clinical Neuroscience, Karolinska Institutet, Stockholm, Sweden.ORCID 0000-0003-3269-7482
Tim SprengerDepartment of Clinical Neuroscience, Karolinska Institutet, Stockholm, Sweden.
Enrico AvventiDepartment of Clinical Neuroscience, Karolinska Institutet, Stockholm, Sweden.ORCID 0000-0003-1000-6134
Ola NorbeckDepartment of Clinical Neuroscience, Karolinska Institutet, Stockholm, Sweden.ORCID 0000-0002-0518-5206
Stefan L GlimbergTracInnovations, Ballerup, Denmark.
Oline V OlesenTracInnovations, Ballerup, Denmark.
Stefan SkareDepartment of Clinical Neuroscience, Karolinska Institutet, Stockholm, Sweden.ORCID 0000-0001-5403-2153
Karolinska University Hospital · SECenter for Clinical & Basic Research · DKKarolinska Institutet · SEGeneral Electric (Spain) · ES

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeTo enable motion-robust diffusion weighted imaging of the brain using well-established imaging techniques.

methodsAn optical markerless tracking system was used to estimate and correct for rigid body motion of the head in real time during scanning. The imaging coordinate system was updated before each excitation pulse in a single-shot EPI sequence accelerated by GRAPPA with motion-robust calibration. Full Fourier imaging was used to reduce effects of motion during diffusion encoding. Subjects were imaged while performing prescribed motion patterns, each repeated with prospective motion correction on and off.

resultsProspective motion correction with dynamic ghost correction enabled high quality DWI in the presence of fast and continuous motion within a 10° range. Images acquired without motion were not degraded by the prospective correction. Calculated diffusion tensors tolerated the motion well, but ADC values were slightly increased.

conclusionsProspective correction by markerless optical tracking minimizes patient interaction and appears to be well suited for EPI-based DWI of patient groups unable to remain still including those who are not compliant with markers.

Indexed as

Diffusion Magnetic Resonance ImagingEcho-Planar ImagingArtifactsBrainHumansMotionProspective Studiesdiffusion tensor imagingdiffusion weighted MRIecho-planar imagingmarkerless motion trackingreal-time movement correction

Identifiers

PMID32989859
PMCPMC7756594
OpenAlexW3091715635

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.