Evidence map›Paper›PMID 32985310›Full record

ArticleThe Journal of international medical research2020

Performance characteristics of an instrument-free point-of-care CD4 test (VISITECT®CD4) for use in resource-limited settings.

Vairamohan Vidhyavathi, Hussain Syed Iqbal, Kannaiyan Kanthamani, Sunil Suhas Solomon, Paneerselvam Nandagopal, Stanley Luchters, Minh D Pham, James Forbes, Suzanne M Crowe, Christopher McMurran and 2 more

Open access · goldAbstract read
In one paragraph

Article in The Journal of international medical research, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
0.4field-weighted citation impact, top 39% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 5 citations in OpenAlex.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

12 authors at 3 institutions in 5 countries.

Vairamohan VidhyavathiYRG Centre for AIDS Research and Education, Voluntary Health Services, Chennai, India.
Hussain Syed IqbalYRG Centre for AIDS Research and Education, Voluntary Health Services, Chennai, India.
Kannaiyan KanthamaniYRG Centre for AIDS Research and Education, Voluntary Health Services, Chennai, India.
Sunil Suhas SolomonYRG Centre for AIDS Research and Education, Voluntary Health Services, Chennai, India.
Paneerselvam NandagopalYRG Centre for AIDS Research and Education, Voluntary Health Services, Chennai, India.
Stanley LuchtersDepartment of Population Health, Aga Khan University, Nairobi, Kenya.
Minh D PhamBurnet Institute, Melbourne, Australia.
James ForbesOmega Diagnostics, Scotland, UK.
Suzanne M CroweBurnet Institute, Melbourne, Australia.
Christopher McMurranOmega Diagnostics, Scotland, UK.
David A AndersonBurnet Institute, Melbourne, Australia.
Pachamuthu BalakrishnanYRG Centre for AIDS Research and Education, Voluntary Health Services, Chennai, India.ORCID https://orcid.org/0000-0002-1710-9155
Voluntary Health Services Hospital · INBurnet Institute · AUJohns Hopkins University · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectiveCD4+ T lymphocyte count remains the most common biomarker of immune status and disease progression in human immunodeficiency virus (HIV)-positive individuals. VISITECT®CD4 is an instrument-free, low-cost point-of-care CD4 test with a cut-off of 350 CD4 cells/μL. This study aimed to evaluate VISITECT®CD4 test's diagnostic accuracy.

methodsTwo hundred HIV-positive patients attending a tertiary HIV centre in South India were recruited. Patients provided venous blood for reference and VISITECT®CD4 tests. An additional finger-prick blood sample was obtained for VISITECT®CD4. VISITECT®CD4's diagnostic performance in identifying individuals with CD4 counts ≤350 cells/μL was assessed by calculating sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) taking flow cytometry as the reference.

resultsThe overall agreement between VISITECT®CD4 and flow cytometry was 89.5% using venous blood and 81.5% using finger-prick blood. VISITECT®CD4 showed better performance using venous blood [sensitivity: 96.6% (95% confidence interval: 92.1%-98.9%), specificity: 70.9% (57.1%-82.4%), PPV: 89.7% (83.9%-94.0%) and NPV: 88.6% (75.4%-96.2%)] than using finger-prick blood [sensitivity: 84.8% (77.9%-90.2%), specificity: 72.7% (59.0%-83.9%), PPV: 89.1% (82.7%-93.8%) and NPV: 64.5% (51.3%-76.3%)].

conclusionVISITECT®CD4 performed well using venous blood, demonstrating its potential utility in decentralization of CD4 testing services in resource-constrained settings.

Indexed as

HIV InfectionsPoint-of-Care SystemsCD4 Lymphocyte CountCD4-Positive T-LymphocytesFlow CytometryHumansIndiaSensitivity and SpecificityCD4+ countHIV/AIDSHIV monitoring assaypoint-of-care diagnosticsresource-limited settingsVISITECT®CD4 test

Identifiers

PMID32985310
PMCPMC7536493
OpenAlexW3087847224

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked