ArticleBMJ open2020

Patch validation: an observational study protocol for the evaluation of a multisignal wearable sensor in patients during anaesthesia and in the postanaesthesia care unit.

Morgan Le Guen, Pierre Squara, Sabrina Ma, Shérifa Adjavon, Bernard Trillat, Messaouda Merzoug, Philippe Aegerter, Marc Fischler

Registry-linked trialOpen access · goldAbstract readClinical Trial Protocol
In one paragraph

Article in BMJ open, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT04344093 (Validation of a Connected Patch, an Alternative to Conventional Monitoring), which is not on this map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
0.6field-weighted citation impact, top 30% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

NCT04344093 nawithdrawnnot on this map

Validation of a Connected Patch, an Alternative to Conventional Monitoring

TypeinterventionalSponsorCMC Ambroise ParéRan2020 to 2022Enrolled0ConditionsAnesthesiaArmsPatch validation
3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 3 citations in OpenAlex.

  1. One small wearable, one giant leap for patient safety?Journal of clinical monitoring and computing · 2022
    Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 3 institutions in 2 countries.

Morgan Le GuenDepartment of Anesthesiology, Hôpital Foch, Suresnes, France.
Pierre SquaraICU, Clinique Ambroise Paré, Neuilly-sur-Seine, France.ORCID 0000-0003-1381-6773
Sabrina MaDepartment of Anesthesiology, Hôpital Foch, Suresnes, France.
Shérifa AdjavonDepartment of Anesthesiology, Hôpital Foch, Suresnes, France.
Bernard TrillatDepartment of Information Systems, Hôpital Foch, Suresnes, France.
Messaouda MerzougResearch Unit, Clinique Ambroise Paré, Neuilly-sur-Seine, France.
Philippe AegerterMethodology Unit, GIRCI-IdF, Paris, France.
Marc FischlerDepartment of Anesthesiology, Hôpital Foch, Suresnes, France m.fischler@orange.fr.ORCID 0000-0003-0729-5430
Hôpital Foch · FRClinique Ambroise Paré · FRInserm · FR

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionExcept for operating rooms, postanaesthesia care units and intensive care units, where the monitoring of vital signs is continuous, intermittent care is standard practice. However, at a time when only the patients with the most serious conditions are hospitalised and only a fraction of these patients are in intensive care units, this type of monitoring is no longer sufficient. Wireless monitoring has been proposed, but it requires rigorous validation. The aim of this observational study is to compare vital signs obtained from a precordial patch sensor to those obtained with conventional monitoring. METHODS AND ANALYSIS: This patch validation trial will be an observational, prospective, single-centre open study of 115 anaesthetised adult patients monitored with both a wireless sensor (myAngel VitalSigns, Devinnova, Montpellier, France) and a standard bedside monitor (Carescape Monitor B850, GE Healthcare, Chicago, Illinois). Both sensors will be used to record peripheral oxygen saturation, respiratory rate, heart rate, body temperature and blood pressure (systolic and diastolic). The main objective will be to assess the degree of agreement between the two systems during the patients' stay in the postanaesthesia care unit, both at the raw signal level and at the clinical parameter level. The secondary objectives will be to assess the same performance under anaesthesia, the frequency of missing data or artefacts, the diagnostic performance of the systems, the influence of patients' characteristics on agreement between the two systems, the adverse events and the acceptability of the patch to patients. Bland-Altman plots will be used in the main analysis to detect discrepancies and estimate the limits of agreement. ETHICS AND DISSEMINATION: Ethics approval was obtained from the Ethical Committee (Toulouse, France) on 10 April 2020. We are not yet recruiting subjects for this study. The results will be submitted for publication in peer-reviewed journals. TRIAL REGISTRATION NUMBER: NCT04344093.

Indexed as

AnesthesiaWearable Electronic DevicesAdultChicagoFranceHumansIllinoisObservational Studies as TopicProspective Studiesadult anaesthesiaadult intensive & critical caresurgerytelemedicine

Identifiers

PMID32978206
PMCPMC7520837
OpenAlexW3088352664

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.