Evidence map›Paper›PMID 32953242›Full record

ArticleTranslational vision science & technology2020

Robotic Retinal Surgery Impacts on Scleral Forces: In Vivo Study.

Müller G Urias, Niravkumar Patel, Ali Ebrahimi, Iulian Iordachita, Peter L Gehlbach

Open access · goldAbstract read
In one paragraph

Article in Translational vision science & technology, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed, 1 pooled it
0.9field-weighted citation impact, top 22% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 1 synthesis or guideline pooled it, 12 citations in OpenAlex.

  1. Pooled it
  2. Article
  3. Review
  4. Review
  5. Article
  6. Autonomous Needle Navigation in Retinal Microsurgery: Evaluation inIEEE International Conference on Robotics and Automation : ICRA : [proceedings]. IEEE International Conference on Robotics and Automation
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 3 institutions in 2 countries.

Müller G UriasWilmer Eye Institute, Johns Hopkins Hospital, Baltimore, MD, USA.
Niravkumar PatelLaboratory for Computational Sensing and Robotics, Johns Hopkins University, Baltimore, MD.
Ali EbrahimiLaboratory for Computational Sensing and Robotics, Johns Hopkins University, Baltimore, MD.
Iulian IordachitaLaboratory for Computational Sensing and Robotics, Johns Hopkins University, Baltimore, MD.
Peter L GehlbachWilmer Eye Institute, Johns Hopkins Hospital, Baltimore, MD, USA.
Johns Hopkins University · USJohns Hopkins Hospital · USUniversidade Federal de São Paulo · BR

Funding

Enabling Technology for Safe Robot-assisted Surgical MicromanipulationR01EB023943 · NIBIB · JOHNS HOPKINS UNIVERSITY · PI GEHLBACH, PETER LOUIS, IORDACHITA, IULIAN IOAN · 2017 to 2024
$2.3M
NIBIB NIH HHS R01 EB023943
6 · The paper itself

Abstract

Purpose: This study aims to map force interaction between instrument and sclera of in vivo rabbits during retinal procedures, and verify if a robotic active force control could prevent unwanted increase of forces on the sclera. Methods: Experiments consisted in the performance of intraocular movements of a force sensing instrument, adjacent to the retinal surface, in radial directions, from the center to the periphery and back, and compared manual manipulations with robotic assistance and also robotic assistance with an active force control. This protocol was approved by the Animal Use and Ethical Committee and experiments were according to ARVO Statement of Animal Use. Results: Mean forces using manual manipulations were 115 ± 51 mN. Using robotic assistance, mean forces were 118 ± 49 mN. Using an active force control method, overall mean forces reduced to 69 ± 15, with a statistical difference compared with other methods ( Conclusions: Results validate that the introduction of robotic assistance might increase the dynamic interactions between instrument and sclera, and the addition of an active force control method reduces the forces at levels lower than manual manipulations. Translational Relevance: All marketing benefits from extreme accuracy and stability from robots, however, redundancy of safety mechanisms during intraocular manipulations, especially on force control and surgical awareness, would allow all utility of robotic assistance in ophthalmology.

Indexed as

RoboticsRobotic Surgical ProceduresAnimalsMicrosurgeryRabbitsRetinaScleramicrosurgeryretinarobotic surgical procedures

Identifiers

PMID32953242
PMCPMC7476674
OpenAlexW3081920524

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.