Biomechanical Stability of Primary and Revision Sacroiliac Joint Fusion Devices: A Cadaveric Study.

Jake Carbone, Isaac Swink, Thomas Muzzonigro, Daniel Diehl, Michael Oh, Derek P Lindsey, Scott Yerby, Boyle C Cheng

Open access · goldAbstract read
In one paragraph

Article in Global spine journal, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
0.1field-weighted citation impact, top 50% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 4 citations in OpenAlex.

  1. Article
  2. Effect of revision sacroiliac joint fusion on unresolved pain and disability: a retrospective cohort study.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2024
    Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 2 institutions in 1 country.

Jake Carbone6596Allegheny Health Network, Pittsburgh, PA, USA.
Isaac Swink6596Allegheny Health Network, Pittsburgh, PA, USA.ORCID https://orcid.org/0000-0003-2794-0847
Thomas Muzzonigro6596Allegheny Health Network, Pittsburgh, PA, USA.
Daniel Diehl6596Allegheny Health Network, Pittsburgh, PA, USA.
Michael Oh6596Allegheny Health Network, Pittsburgh, PA, USA.
Derek P Lindsey480623SI-BONE, Inc. Santa Clara, CA, USA.ORCID https://orcid.org/0000-0003-2699-8190
Scott Yerby480623SI-BONE, Inc. Santa Clara, CA, USA.
Boyle C Cheng6596Allegheny Health Network, Pittsburgh, PA, USA.
Allegheny Health Network · USSI-Bone (United States) · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

STUDY

designAn in vitro biomechanics study.

objectiveTo evaluate the efficacy of triangular titanium implants in providing mechanical stabilization to a sacroiliac joint with primary and revision sized implants.

methodsTen lumbopelvic cadaveric specimens were tested in 4 stages: intact, pubic symphysis sectioned, primary, and simulated revision. Primary treatment was performed using 3 laterally placed triangular titanium implants. To simulate revision conditions before and after bone ingrowth and ongrowth on the implants, 7.5-mm and 10.75-mm implants were randomly assigned to one side of each specimen during the simulated revision stage. A 6 degrees of freedom spinal loading frame was used to load specimens in 4 directions: flexion extension, lateral bending, axial torsion, and axial compression. Biomechanical evaluation was based on measures of sacroiliac joint rotational and translational motion.

resultsBoth primary and revision implants showed the ability to reduce translational motion to a level significantly lower than the intact condition when loaded in axial compression. Simulated revision conditions showed no statistically significant differences compared with the primary implant condition, with the exception of flexion-extension range of motion where motions associated with the revised condition were significantly lower. Comparison of rotational and translation motions associated with the 7.5- and 10.75-mm implants showed no significant differences between the treatment conditions.

conclusionsThese results indicate that implantation of laterally placed triangular titanium implants significantly reduces the motion of a sacroiliac joint using either the primary and revision sized implants. No statistically significant differences were detected when comparing the efficacy of primary, 7.5-mm revision, or 10.75-mm revision implants.

Indexed as

biomechanicsfusionrevisionsacroiliacspine

Identifiers

PMID32935574
PMCPMC8965313
OpenAlexW3087712860

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.