Evidence map›Paper›PMID 32928018›Full record

Observational studyActa orthopaedica2021

Custom-made 3D-printed cup-cage implants for complex acetabular revisions: evaluation of pre-planned versus achieved positioning and 1-year migration data in 10 patients.

Vasileios Zampelis, Gunnar Flivik

Abstract readObservational Study
In one paragraph

Observational study in Acta orthopaedica, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 18 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
18citing papers in PubMed, 1 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

18 citing papers in PubMed, 1 synthesis or guideline pooled it.

  1. Pooled it
  2. Trial
  3. Article
  4. Article
  5. Review
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Article
  12. Review
  13. Article
  14. Patient-specific 3D printing to replace components of a rib-to pelvis "Eiffel Tower" vertebral expanding prosthetic titanium rib system in an infant: a case report.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2023
    Article
  15. Article
  16. Article
  17. Article
  18. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Vasileios ZampelisDepartment of Orthopedics, Skane University Hospital, Clinical Sciences, Lund University, Lund, Sweden.
Gunnar FlivikDepartment of Orthopedics, Skane University Hospital, Clinical Sciences, Lund University, Lund, Sweden.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background and purpose - The use of custom-made 3D-printed titanium implants for the reconstruction of large acetabular defects has been successively introduced in the last decade. In an observational cohort study we evaluated the agreement between preoperatively planned and actually achieved cup-cage position as well as 1-year follow-up migration of the cup-cage component.Patients and methods - 10 patients with Paprosky III defects underwent revision surgery using a custom-made 3D-printed cup-cage. The position of the implant on postoperative CT scan was compared with the preoperative plan and the postoperative CT scan was compared with the 1-year follow-up CT scan.Results - There was a median deviation in postoperative position versus planned in inclination of 3.6° (IQR 1.0-5.4), in anteversion of -2.8° (IQR -7.5 to 1.2), and in rotation of -1.2° (IQR -3.3 to 0.0). The median deviation in position of the center of rotation (COR) was -0.5 mm (IQR 2.9 to 0.7) in the anteroposterior (AP) plane, -0.6 mm (IQR -1.8 to -0.1) in the mediolateral (ML) plane, and 1.1 mm (IQR -1.6 to 2.8) in the superoinferior (SI) plane. The migration between postoperative and 1-year follow-up caused a mean change in inclination of 0.04° (IQR -0.06 to 0.09), in anteversion of -0.13° (IQR -0.23 to -0.06), and in rotation of 0.05° (IQR -0.46 to 1.4). The migration of COR was -0.08 mm (IQR -0.18 to -0.04) in the AP plane, 0.14 mm (IQR -0.08 to 0.22) in the ML plane, and 0.06 mm (IQR -0.02 to 0.35) in the SI plane. There was no re-revision.Interpretation - The early results show good agreement between planned and achieved cup-cage position and small measured migration values of the cup-cage component at the 1-year follow-up.

Indexed as

Hip ProsthesisProsthesis DesignProsthesis FittingAcetabulumAdultAgedAged, 80 and overArthroplasty, Replacement, HipCohort StudiesFemaleHumansMaleMiddle AgedPrinting, Three-DimensionalReoperationTitaniumTitanium

Identifiers

PMID32928018
PMCPMC7919918

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.