From a PMT-based to a SiPM-based PET system: a study to define matched acquisition/reconstruction parameters and NEMA performance of the Biograph Vision 450.
Article in EJNMMI physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 14 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
14 citing papers in PubMed.
- Development of a Prototype AWSM-PET Device for Augmented Whole-Body PET Imaging and Initial Human Study.Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2026Article
- A data-driven approach for spatial alignment of PET add-on detectors with sub-millimeter accuracy.Physics in medicine and biology · 2025Article
- Consistency Analysis of Centiloid Values Across Three Commercial Software Platforms for Amyloid PET Quantification.Diagnostics (Basel, Switzerland) · 2025Article
- PET/CT acquisition and processing protocols in the Netherlands.EJNMMI physics · 2025Article
- Review
- Comparison of Digital versus AnalogDiagnostics (Basel, Switzerland) · 2023Article
- Comparative studies of the sensitivities of sparse and full geometries of Total-Body PET scanners built from crystals and plastic scintillators.EJNMMI physics · 2023Article
- PET imaging and quantification of small animals using a clinical SiPM-based camera.EJNMMI physics · 2023Article
- Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra.European journal of nuclear medicine and molecular imaging · 2023Article
- Phantom-based acquisition time and image reconstruction parameter optimisation for oncologic FDG PET/CT examinations using a digital system.BMC cancer · 2022Article
- The effect of modern PET technology and techniques on the EANM paediatric dosage card.European journal of nuclear medicine and molecular imaging · 2022Article
- Yttrium-90 quantitative phantom study using digital photon counting PET.EJNMMI physics · 2021Article
- Performance comparison of a dedicated total breast PET system with a clinical whole-body PET system: a simulation study.Physics in medicine and biology · 2021Article
- Assessment of Image Quality and Lesion Detectability With Digital PET/CT System.Frontiers in medicine · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
Abstract
backgroundThe purpose of this work was to propose an approach based on noise measurement to adapt present clinical acquisition and reconstruction parameters adapted to a PMT-based system (Biograph mCT) to a SiPM-based system (Biograph Vision 450) sharing identical geometrical properties. The NEMA performance (NEMA) of the recently released Biograph Vision 450 PET/CT (Vision) was also derived.
methodsAll measurements were conducted on Vision and Biograph mCT with TrueV (mCT). A full NEMA-based performance was derived for Vision only. The adaptation of acquisition and reconstruction parameters from mCT to Vision was done using the NEMA image quality phantom. The noise level reached using mCT was set as the reference value for six different numbers of net true coincidences. The noise level computed using Vision was matched to the reference noise level (within 0.01%) using a different reconstruction set-up to determine the potential reduction of count numbers for the same noise level.
resultsVision sensitivity was 9.1 kcps/MBq for a timing resolution of 213 ps at 5.3 kBq/mL. The NEMA-based CR for the 10-mm sphere was better than 75% regardless the reconstruction set-up studied. The mCT reference noise properties could be achieved using Vision with a scan time reduction (STR) of 1.34 with four iterations and a 440 × 440 matrix size (or STR = 1.89 with a 220 × 220 matrix size) together with a 3D CR improvement of 53% for the 10-mm sphere (24% using 220 × 220).
conclusionThe Vision exhibited improved NEMA performances compared to mCT. Using the proposed approach, the time acquisition could be divided by almost two, while keeping the same noise properties as that of mCT with a marked improvement of contrast recovery.