ArticleThe journal of pediatric pharmacology and therapeutics : JPPT : the official journal of PPAG2020
Incidence and Causes of Infusion Alarms in a Neonatal and Pediatric Intensive Care Unit: A Prospective Pilot Study.
Article in The journal of pediatric pharmacology and therapeutics : JPPT : the official journal of PPAG, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed, 13 citations in OpenAlex.
- Article
- Good Clinical Practice Guidance for Line Materials, Filtration, and Light Protection in Intravenous Medication Administration: Modified Delphi Consensus Study.JMIR human factors · 2026Article
- Simulated Y-Site Compatibility of Calcium Gluconate and Benzylpenicillin With Maintenance Fluid Used in Neonatal Intensive Care Unit.Clinical and translational science · 2026Article
- Crying wolf, alarm safety and management in paediatrics: A scoping review.Journal of advanced nursing · 2025Article
- AF among Nurses Working in Neonatal and Paediatric Intensive Care Units: A Cross-Sectional Study.Healthcare (Basel, Switzerland) · 2024Article
- Article
- Critical Care Nurses' Knowledge of Correct Line Types for Administration of Common Intravenous Medications: Assessment and Intervention Study.JMIR formative research · 2022Article
- Quantifying the Impact of Infusion Alerts and Alarms on Nursing Workflows: A Retrospective Analysis.Applied clinical informatics · 2021Article
- Intravenous Smart Pump Alarms in Everyday Hospital Use: A Real-world Descriptive Analysis.Dimensions of critical care nursing : DCCNArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesTo evaluate the incidence and causes of infusion alarms in a NICU/PICU setting.
methodsWe conducted a 90-day prospective analysis of event logs downloaded daily from infusion pumps (syringe and volumetric pumps). The details about conditions surrounding alarm events were described daily by bedside nurses on a standardized form. The occlusion pressure alarm was set at 300 mm Hg on each device.
resultsForty-one pediatric patients including 12 neonates, mean weight 11.0 ± 11.3 kg (minimum-maximum, 0.48-50), were included for a total infusion time of 2164 hours. Eight hundred forty-three infusion alarms were documented (220 [26.1%] occlusion; 273 [32.4%] infusion completed; 324 [38.4%] door open/syringe disengagement; 26 [3.1%] air-in-line) resulting in an incidence of 4.7 infusion (1.2 occlusion) alarms per patient per day.Detailed conditions surrounding occlusion alarm events were documented in only 22.7% (50/220) of the cases. Of these, 36% (18/50) were related to closed or clamped lines, 4% (2/50) to syringe change, 16% (8/50) to drug injection, and 8% (4/50) to patient-related factors. The remaining 36% (18/50) occurred without any apparent external cause during ongoing infusion, among these drug incompatibilities were a potential cause for 12 events.
conclusionAlarms from infusion pumps were frequent in the NICU/PICU setting, a quarter of them resulting from line occlusion. Other than well-known triggers (mechanical and patient factors), drug incompatibilities were identified as a potential cause for occlusion alarms in this pilot study.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.