Commissioning and quality assurance of a novel solution for respiratory-gated PBS proton therapy based on optical tracking of surface markers.
Article in Zeitschrift fur medizinische Physik, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. An erratum has been issued. Cited by 10 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
10 citing papers in PubMed, 20 citations in OpenAlex.
- Interplay effect evaluation on lung and esophageal cancer patients treated with compact proton system in free breathing.Physics and imaging in radiation oncology · 2026Article
- Commissioning of respiratory-gated 4D dynamic dose calculations for various gating widths without spot timestamp in proton pencil beam scanning.Medical physics · 2025Article
- Log file-based quality assurance method for respiratory gating system.Journal of applied clinical medical physics · 2025Article
- Resolving signal drift in the wall-mounted camera of the RGSC system.Journal of applied clinical medical physics · 2024Article
- Characteristics of breathing-adapted gating using surface guidance for use in particle therapy: A phantom-based end-to-end test from CT simulation to dose delivery.Journal of applied clinical medical physics · 2024Article
- Emerging technologies for cancer therapy using accelerated particles.Progress in particle and nuclear physics · 2023Article
- A survey of practice patterns for real-time intrafractional motion-management in particle therapy.Physics and imaging in radiation oncology · 2023Article
- Advances and potential of optical surface imaging in radiotherapy.Physics in medicine and biology · 2022Review
- Technical assessment of the NDI Polaris Vega optical tracking system.Radiation oncology (London, England) · 2021Article
- Development and Performance Evaluation of Wearable Respiratory Self-Training System Using Patch Type Magnetic Sensor.Frontiers in oncology · 2021Article
Corrections and comments
- Erratum issued
Authors and funding
19 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
We present the commissioning and quality assurance of our clinical protocol for respiratory gating in pencil beam scanning proton therapy for cancer patients with moving targets. In a novel approach, optical tracking has been integrated in the therapy workflow and used to monitor respiratory motion from multiple surrogates, applied on the patients' chest. The gating system was tested under a variety of experimental conditions, specific to proton therapy, to evaluate reaction time and reproducibility of dose delivery control. The system proved to be precise in the application of beam gating and allowed the mitigation of dose distortions even for large (1.4cm) motion amplitudes, provided that adequate treatment windows were selected. The total delivered dose was not affected by the use of gating, with measured integral error within 0.15cGy. Analysing high-resolution images of proton transmission, we observed negligible discrepancies in the geometric location of the dose as a function of the treatment window, with gamma pass rate greater than 95% (2%/2mm) compared to stationary conditions. Similarly, pass rate for the latter metric at the 3%/3mm level was observed above 97% for clinical treatment fields, limiting residual movement to 3mm at end-exhale. These results were confirmed in realistic clinical conditions using an anthropomorphic breathing phantom, reporting a similarly high 3%/3mm pass rate, above 98% and 94%, for regular and irregular breathing, respectively. Finally, early results from periodic QA tests of the optical tracker have shown a reliable system, with small variance observed in static and dynamic measurements.