ArticleSurgical endoscopy2021
Electromagnetic interference on cardiac pacemakers and implantable cardioverter defibrillators during endoscopy as reported to the US Federal Drug Administration.
Article in Surgical endoscopy, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed, 17 citations in OpenAlex.
- Sedation reversal trends at outpatient ambulatory endoscopic centerWorld journal of gastrointestinal endoscopy · 2024Trial
- Radiotherapy in Patients with Cardiac Implantable Devices: A Single-Centre Retrospective Observational Analysis of Local Guidelines.Journal of clinical medicine · 2026Article
- Home-Based Self-Management After Permanent Pacemaker Implantation: What Should Patients Know.Patient preference and adherence · 2025Review
- Review
- A Systematic Review and Meta-analysis of the Prevalence and Risk Factors in Cardiac Implantable Electronic Device Malfunction.Journal of the Saudi Heart Association · 2023Review
- EHRA consensus on prevention and management of interference due to medical procedures in patients with cardiac implantable electronic devices.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2022Article
- Effects of medically generated electromagnetic interference from medical devices on cardiac implantable electronic devices: A review.World journal of cardiology · 2022Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionMore than 3 million patients have a cardiac implanted electronic device (CIED) such as a pacemaker or implanted cardioverter-defibrillator in the USA. These devices are susceptible to electromagnetic interference (EMI) leading to malfunction and injury. Radiofrequency energy, the most common modality for obtaining hemostasis during endoscopy, is the most common source of EMI. Few studies have evaluated the effect of endoscopic radiofrequency energy on CIEDs. We aim to characterize CIED dysfunction related to endoscopic procedures. We hypothesize that EMI from endoscopic energy can result in patient injury.
methodsWe queried the Manufacturer and User Facility Device Experience (MAUDE) database for CIED dysfunction related to electrosurgical devices over a 10-year period (2009-2019). CIED dysfunction events were identified using seven problem codes (malfunction, electromagnetic interference, ambient noise, pacing problem, over-sensing, inappropriate shock, defibrillation). These were cross-referenced for the terms "cautery, electrocautery, endoscopy, esophagus, colonoscopy, colon, esophagoscopy, and esophagogastroduodenoscopy." Reports were individually reviewed to confirm and characterize CIED malfunction due to an endoscopic procedure.
resultsA search for CIED dysfunction resulted in 43,759 reports. Three hundred and eleven reports (0.7%) were associated with electrocautery, and of these, 45 reports (14.5%) included endoscopy. Ten reports involving endoscopy (22%) specified upper (3, 7%) or lower (7, 16%) endoscopy while the remainder were non-specific. Twenty-six of reports involving endoscopy (58%) suffered injury because of CIED dysfunction: Of these, 17 (65%) received inappropriate shocks, 5 (19%) had pacing inhibition with bradycardia or asystole, 3 (12%) had CIED damage requiring explant and replacement, and 1 (4%) patient suffered ventricular tachycardia requiring hospital admission.
conclusionThe use of energy during endoscopy can cause dysfunction of CIEDs. This most commonly results in inappropriate defibrillation, symptomatic bradycardia, and asystole. Patients with CIEDs undergoing endoscopy should undergo pre- and post-procedure device interrogation and re-programming to avoid patient injury.