A method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease.
Article in Biochemia medica, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 3 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionHyperkalaemia is a common electrolyte disorder that may cause life-threatening cardiac arrythmias. We aimed to determine the agreement of potassium concentrations between GEM premier 3500 point-of-care blood gas analyser (POC-BGA) and Roche Cobas 6000 c501 auto-analyser in patients with hyperkalaemia.
methodsA prospective, cross-sectional study of all consecutive adult patients referred to the Renal Unit with a serum potassium concentration ≥ 5.5 mmol/L was performed. A total of 59 paired venous blood samples were included in the final statistical analysis. Passing-Bablok regression and Bland Altman analysis were used to compare the two methods.
resultsThe median laboratory auto-analyser potassium concentration was 6.1 (5.9-7.1) mmol/L as compared to the POC-BGA potassium concentration of 5.7 (5.5-6.8) mmol/L with a mean difference of - 0.43 mmol/L and 95% upper and lower limits of agreement of 0.35 mmol/L and - 1.21 mmol/L, respectively. Regression analysis revealed proportional systematic error. Test for linearity did not indicate significant deviation (P = 0.297).
conclusionAlthough regression analysis indicated proportional systematic error, on Bland Altman analysis, the mean difference appeared to remain relatively constant across the potassium range that was evaluated. Therefore, in patients presenting to the emergency department with a clinical suspicion of hyperkalaemia, POC-BGA potassium concentrations may be considered a surrogate for laboratory auto-analyser measurements once clinicians have been cautioned about this difference.