Evidence map›Paper›PMID 32764313›Full record

ArticleBioengineering (Basel, Switzerland)2020

Development of a Novel Medical Device for Mucositis and Peri-Implantitis Treatment.

Gloria Cosoli, Lorenzo Scalise, Alfredo De Leo, Paola Russo, Gerardo Tricarico, Enrico Primo Tomasini, Graziano Cerri

Open access · goldAbstract read
In one paragraph

Article in Bioengineering (Basel, Switzerland), 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
0.7field-weighted citation impact, top 31% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 3 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 2 institutions in 1 country.

Gloria CosoliDepartment of Industrial Engineering and Mathematical Sciences, Università Politecnica delle Marche, 60131 Ancona, Italy.ORCID 0000-0001-7982-208X
Lorenzo ScaliseDepartment of Industrial Engineering and Mathematical Sciences, Università Politecnica delle Marche, 60131 Ancona, Italy.ORCID 0000-0002-4728-3736
Alfredo De LeoDepartment of Information Engineering, Università Politecnica delle Marche, 60131 Ancona, Italy.
Paola RussoDepartment of Information Engineering, Università Politecnica delle Marche, 60131 Ancona, Italy.
Gerardo TricaricoDepartment of Health Sciences, Università del Piemonte Orientale, 28100 Novara, Italy.
Enrico Primo TomasiniDepartment of Industrial Engineering and Mathematical Sciences, Università Politecnica delle Marche, 60131 Ancona, Italy.
Graziano CerriDepartment of Information Engineering, Università Politecnica delle Marche, 60131 Ancona, Italy.
Marche Polytechnic University · ITUniversità degli Studi del Piemonte Orientale “Amedeo Avogadro” · IT

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

In spite of all the developments in dental implantology techniques, peri-implant diseases are frequent (prevalence up to 80% and 56% of subjects for mucositis and peri-implantitis, respectively) and there is an urgency for an effective treatment strategy. This paper presents an innovative electromedical device for the electromagnetic treatment of mucositis and peri-implantitis diseases. This device is also equipped with a measurement part for bioimpedance, which reflects the health conditions of a tissue, thus allowing clinicians to objectively detect impaired areas and to monitor the severity of the disease, evaluate the treatment efficacy, and adjust it accordingly. The design of the device was realized considering literature data, clinical evidence, numerical simulation results, and electromagnetic compatibility (EMC) pre-compliance tests, involving both clinicians and engineers, to better understand all the needs and translate them into design requirements. The reported system is being tested in more than 50 dental offices since 2019, providing efficient treatments for mucositis and peri-implantitis, with success rates of approximately 98% and 80%, respectively.

Indexed as

bioimpedancedental implantsdentistryelectromagnetic compatibilityimmunity testingmedical devicemucositisperi-implant diseasesperi-implantitisradio frequency

Identifiers

PMID32764313
PMCPMC7552606
OpenAlexW3047123754

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked