ArticleJMIR human factors2020
Twelve-Month Review of Infusion Pump Near-Miss Medication and Dose Selection Errors and User-Initiated "Good Save" Corrections: Retrospective Study.
Article in JMIR human factors, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
6 citing papers in PubMed.
- Assessing Medication CDS Usability: Pilot Results from 10 Outpatient Clinics.Applied clinical informatics · 2025Article
- The Introduction of Robotics to an Outpatient Dispensing and Medication Management Process in Saudi Arabia: Retrospective Review of a Pharmacy-led Multidisciplinary Six Sigma Performance Improvement Project.JMIR human factors · 2022Article
- The impact of drug error reduction software on preventing harmful adverse drug events in England: a retrospective database study.BMJ open quality · 2022Article
- Critical Care Nurses' Knowledge of Correct Line Types for Administration of Common Intravenous Medications: Assessment and Intervention Study.JMIR formative research · 2022Article
- Dose error reduction software in medication safety risk management - optimising the smart infusion pump dosing limits in neonatal intensive care unit prior to implementation.BMC pediatrics · 2022Article
- Evaluating An Automated Compounding Workflow Software for Safety and Efficiency: Implementation Study.JMIR human factors · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThere is a paucity of quantitative evidence in the current literature on the incidence of wrong medication and wrong dose administration of intravenous medications by clinicians. The difficulties of obtaining reliable data are related to the fact that at this stage of the medication administration chain, detection of errors is extremely difficult. Smart pump medication library logs and their reporting software record medication and dose selections made by users, as well as cancellations of selections and the time between these actions. Analysis of these data adds quantitative data to the detection of these kinds of errors.
objectiveWe aimed to establish, in a reproducible and reliable study, baseline data to show how metrics in the set-up and programming phase of intravenous medication administration can be produced from medication library near-miss error reports from infusion pumps.
methodsWe performed a 12-month retrospective review of medication library reports from infusion pumps from across a facility to obtain metrics on the set-up phase of intravenous medication administration. Cancelled infusions and resolutions of all infusion alerts by users were analyzed. Decision times of clinicians were calculated from the time-date stamps of the pumps' logs.
resultsIncorrect medication selections represented 3.45% (10,017/290,807) of all medication library alerts and 22.40% (10,017/44,721) of all cancelled infusions. Of these cancelled medications, all high-risk medications, oncology medications, and all intravenous medications delivered to pediatric patients and neonates required a two-nurse check according to the local policy. Wrong dose selection was responsible for 2.93% (8533/290,807) of all alarms and 19.08% (8533/44,721) of infusion cancellations. Average error recognition to cancellation and correction times were 27.00 s (SD 22.25) for medication error correction and 26.52 s (SD 24.71) for dose correction. The mean character count of medications corrected from initial lookalike-soundalike selection errors was 13.04, with a heavier distribution toward higher character counts. The position of the word/phrase error was spread among name beginning (6991/10,017, 69.79%), middle (2144/10,017, 21.40%), and end (882/10,017, 8.80%).
conclusionsThe study identified a high number of lookalike-soundalike near miss errors, with cancellation of one medication being rapidly followed by the programming of a second. This phenomenon was largely centered on initial misreadings of the beginning of the medication name, with some incidences of misreading in the middle and end portions of medication nomenclature. The value of an infusion pump showing the entire medication name complete with TALLman lettering on the interface matching that of medication labeling is supported by these findings. The study provides a quantitative appraisal of an area that has been resistant to study and measurement, which is the number of intravenous medication administration errors of wrong medication and wrong dose that occur in clinical settings.